75
Magtrol Model DSP7000 Dynamometer Controller
Chapter 7 – Computer Controlled Operation
OPERA
TION
7.4.1
a
larM
C
oMManDS
Command
Code
Channel 1-
Command
Code
Channel 2
function
explanation
ALA1,#
ALA2,#
Enables or disables air flow
alarm.
Values for # are:
0 = disable
1 = enable
ALC,#
ALC,#
Enables or disables clutch
alarm.
Values for # are:
0 = disable
1 = enable
ALE1,#
ALE2,#
Enables or disables external
alarm.
Values for # are:
0 = disable
1 = enable
ALL1,#
ALL2,#
Enables or disables all alarms. Values for # are:
0 = disable
1 = enable
ALP1,xx.xx
ALP2,xx.xx
Sets power alarm.
Sets maximum power in
kilowatts. Range is 0 to
99,999. Setting is applied to
current channel.
ALS1,xx.xx
ALS2,xx.xx
Sets speed alarm point.
Sets maximum speed in rpm.
Range is 0 to 99,999. Setting
is applied to current channel.
ALT1,xx.xx
ALT2,xx.xx
Sets torque alarm point.
Sets maximum torque in input
units. Range is 0 to 10,000.
Setting is applied to current
channel.
ALW1,#
ALW2,#
Enables or disables water flow
alarm.
Values for # are:
0 = disable
1 = enable
ALR1,#
ALR2,#
Set up the Relay 1 on the IO
card, either channel one or two,
to be used in conjunction with
that channels alarm system.
See section 5.1.1 Alarm Relay
Values for # are:
0 = disable
1 = enable
GP, xx.xx
Set up the global power alarm
on WB/WB or PB/PB tandem
configuration.
Set maximum global power in
Kilowatts. Range is from 0 to
99,999..
GT, xx.xx
Set up the global torque alarm
on WB/WB or PB/PB tandem
configuration.
Set maximum global torque
in N·m. Range is from 0 to
99,999..