7.1.5.4
Basic Calibration Process
1. Turn on the unit with ShIFT button pressed.
2. Send CM1 command. This sets all DAC outputs to 10 volts nominal.
3. Using a volt meter, measure the voltage on the channel.
4. Send the appropriate CG command. See
Section 7.1.5.3 – Calibration Commands
.
5. Repeat steps 3 and 4 until all 3 channels are complete.
6. Send CM2 command. This sets all DAC outputs to -10 volts nominal.
7. Using a volt meter, measure the voltage on the channel.
8. Send the appropriate CG command. See
Section 7.1.5.3 – Calibration Commands
.
9. Repeat steps 7 and 8 until all 3 channels are complete.
10. Send Cx command to save calibration values.