Chapter 3 – Installation/Configuration
SETUP
SETUP
3.3.5
V
oltS
S
caling
The voltage measurement range can be extended by using a potential transformer. Frequency response
will be determined by the characteristics of the transformer used.
3.3.5.1
Hardware Connection
See
Figure 3–12 Current/Potential Transformer Connection
.
3.3.5.2
Software Configuration
To configure the 6510
e
volt scaling for a potential transformer, complete the following steps utilizing
the USER MENU located on the front panel of the unit.
1. Turn on the 6510
e
. See
Section 3.1 – Powering Up the 6510e
.
2. Press ENTER button. The display should appear as shown in
Figure 2–3 Device Setup
Menu
.
3. Press
and
buttons until VoLT SCALING appears in the display.
4. Press ENTER button. The display will appear as follows.
CURRENT
VOLTAGE
POWER/PF
Figure 3–15 Volt Scaling Setup Menu
5. To set volt scaling, press the RMS button under V1 and use the
pq
buttons in the
USER MENU to adjust the value.
6. To turn scaling oN, press the PEAk-CF button under oFF until oN is reached.
7. Press ENTER to exit the Device Setup Menu.
Note:
When volts scaling has been activated, “
*
” will appear in the main
menu next to the volts range as shown in
Figure 3–14 Amp/Volt
Scaling Activated
.