41
Magtrol Model 6510
e
Single-Phase Power Analyzer
Chapter 5 – Computer Controlled Operation
OPERA
TION
OPERA
TION
5.4.2
oE E
xaMPlE
Total
= 43 characters
Output String: (1-42) = measurement value float E notation (ANSI)
Data Position: Ax, Vx, Wx (where x = requested phase)
5.4.2.1
Good Response
\s\s1.83352E-01,\s\s1.02,\s\s1.01\r\n
5.4.2.2
Over-Range Condition
\s\s1.84250E-01,^\s4.01,^\s4.00\r\n
5.4.3
oa/oV/oW/oF E
xaMPlE
Total
= 15 characters
Output String: (1-14) = measurement value float E notation (ANSI)
5.4.3.1
Good Response
\s\s1.02\r\n
5.4.3.2
Over-Range Condition
^\s4.01\r\n
5.5
programming
Note:
Check the manual provided with your software for full
instructions.
5.5.1
d
ata
t
ErMination
c
haractErS
Use the following information to answer the formatting questions asked when installing your GPIb
software. All GPIb data acquisition systems require the use of data termination characters. The
6510
e
uses the GPIb standard termination characters Carriage Return (CR) and Line Feed (LF).
Provide them in that order.
5.5.12
Codes for CR-LF
BASIC
HEX
DEC
CR =
CHR$(13)
OD
13
LF =
CHR$(10)
OA
10