76
(17) Index <400B+(N-1)×80 hex>: Detection Function Setting
Sub
Index
Contents
Size
Access PDO
map
Setting range
(default setting)
0
Number of Entries 1Byte
(U8)
RO
Not
possible
01 hex
(01 hex)
1
Detection
Function Setting
2Byte
(U16)
RW
Not
possible
0000 hex to FFFF hex
0000 hex: SHS (super high
speed)
0001 hex: HS (high speed)
0002 hex: STND
(standard)
0003 hex: GTGA (high
resolution)
0004 hex to FFFF hex
:
Not
used
(FFFF hex)
Sub-index 1 sets the detection function.
* The “Detection Function Setting” function is not available for the MG80-
EC, even if set.
(18) Index <4011+(N-1)×80 hex>: Display Digits Setting
Sub
Index
Contents
Size
Access PDO
map
Setting range
(default setting)
0
Number of Entries 1Byte
(U8)
RO
Not
possible
01 hex
(01 hex)
1
Display Digits
Setting
2Byte
(U16)
RW
Not
possible
0000 hex to FFFF hex
0000 hex: 4 digits
0001 hex: 3 digits
0002 hex: 2 digits
0003 hex: 1 digits
0004 hex to FFFF hex: Not
used
(FFFF hex)
Sub-index 1 sets the display digits.
* The “Display Digits Setting” function is not available for the MG80-EC,