![Magnescale MG80-EC Operating Manual Download Page 72](http://html1.mh-extra.com/html/magnescale/mg80-ec/mg80-ec_operating-manual_3564493072.webp)
70
(6) Index <300A hex>: Measuring Unit Communication Status
8bit
Sub
Index
Contents
Size
Access PDO
map
Setting range
(default setting)
0
Number of Entries
1Byte
(U8)
RO
Not
possible
01 hex
(01 hex)
1
Measuring Unit
Communication
Status 8bit
1Byte
(U8)
RO
Possible 00 hex to 02 hex
Bit 0 = ON:
Communication in
progress
Bit 1 = ON: Number of
connected measuring
units disagreement
(00 hex)
Sub-index 1 indicates the measuring unit communication status.
* “Measuring Unit Communication Status 8bit” is fixed to 00 hex for the
MG80-EC.
(7) Index <300B hex>: Measuring Unit Warning Status
Sub
Index
Contents
Size
Access PDO
map
Setting range
(default setting)
0
Number of
Entries
1Byte
(U8)
RO
Not
possible
01 hex
(01 hex)
1
Measuring
Unit Status
4Byte
(U32)
RO
Possible 0000 0000 hex to 3FFF FFFF hex
Corresponding bit = ON: Warning
occurred
(0000 0000 hex)
Sub-index 1 gives the measuring unit warning status.
* “Measuring Unit Warning Status” is fixed to 0000 0000 hex for the
MG80-EC.