Quickstart guide NANOCOLOR

®

 

UV

/

VIS

 II / VIS II V 1.02 / 08.15

6.2 MN tests

6.2.1 Tube tests

With the spectrophotometer MACHEREY-NAGEL tube tests can be selected in 

different ways. All test tubes are labeled with a barcode. When the start screen 

is active or the device is in measuring mode, the barcode of the inserted test 

tube will be scanned automatically. The instrument detects the tube test and the measur-

ing process is initiated automatically (tube tests which need to be measured against a 

zero solution cannot be measured automatically).
After the measurement has been completed, the measurement result will be displayed 

in the result window. Via 

 and the icons below the result window, different sample 

information can be added. Removal of the cuvette or leaving the menu via the remainder 

of the task bar icons quits the measurement process and stores the measurement result 

in the measurement memory of the device. 

UV/VIS II

Open Mode, 28.10.2014, 10:57

Measure

Zero

0010

Sample author Sample 

place

Comment

49 mg/L Mg

86 mg/L Ca

> 25 °e (29.2)

0444 Hardness Ca / Mg

540 nm

86 mg/L Ca

0.567 E

IQC

Fig. 17: Result

In addition to opening the measurement menu through barcode identification, in the test 

menu it is possible to manually select the tube test from a list through 

 or to manually 

insert the test number through 

 1  2  3 
 4  5  6
 7  8  9

 .

In case of these two options, the measurement will not be triggered automatically after 

insertion of the test tube. In these cases you can initiate the measurement by pressing 

.

Summary of Contents for NANOCOLOR VIS II

Page 1: ...Smart photometry NANOCOLOR VIS II NANOCOLOR UV VIS II Quickstart guide Distributed By...

Page 2: ......

Page 3: ...et up location 10 3 2 Package content 10 4 Outer appearance 11 4 1 Front and side view 11 4 2 Back view 12 5 Operating instruction 13 5 1 Connectivity 13 5 2 Turning on the instrument 13 5 3 Operation...

Page 4: ......

Page 5: ...comfortable and enjoyable handling of the spectrophotometer Thus the spectrophotometer is the ideal laboratory photometer for the entire spectrum of water analysis 1 1 Technical description A halogen...

Page 6: ...More than 200 preprogrammed tests and special methods 100 user defined methods absorbance transmission factor kinetics 2 point calibration scan nephelometric turbidity measurement Turbidity measureme...

Page 7: ...W H 400 440 170 mm 360 400 110 mm Weight 6 5 kg 4 0 kg Warranty 2 years Housing protection type IP20 Protection rating class 1 class 3 Declaration of conformity EMC directive 2014 30 EU EN 61326 1 200...

Page 8: ...power cable or power adapter only 2 2 Biological risk When using the instrument it may be necessary to handle hazardous chemicals or biologically harmful samples Please adhere to the following guideli...

Page 9: ...UV VIS II features a halogen and a deuterium lamp The NANOCOLOR VIS II features a halogen lamp Both get very hot during opera tion Hence there is a risk of severe burns Before exchanging the lamps le...

Page 10: ...re not to damage its contents Remove the instrument and all other parts carefully Check the package inst rument and accessories for visible damages In case a part is damaged please contact your distri...

Page 11: ...Side view UV VIS II a a b c e Fig 3 Front view VIS II f Fig 4 Side view VIS II a Stereo speakers b 10 1 HD touch screen with PCAP c Universal cuvette slot for tube tests of 16 mm OD and rectangular c...

Page 12: ...Back view UV VIS II 7 1 2 3 4 8 6 Fig 6 Back view VIS II 1 USB A interface 2 USB B interface 3 Ethernet LAN interface 4 RS 232 interface 5 Connection for power supply 110 240 V 50 60 Hz 6 Main switch...

Page 13: ...l as laboratory infor mation management systems LIMS Next to the electrical connection Fig 6 8 5 more types of interfaces are available within the NANOCOLOR VIS II as indicated on page 10 1 x SDHC car...

Page 14: ...eter features a 10 1 HD display with a projective capacitive touch screen PCAP The instrument s cover glass is non reflecting It can be easily and conveniently cleaned with the microfiber cloth includ...

Page 15: ...for projective capacitive touch screens Wearing protective gloves does not impair handling the instrument properly In selected menus you can zoom in and out by moving two fingers in opposite direction...

Page 16: ...unctions Home icon By pressing this icon at any position in the menu one is guided back to the start screen see Fig 7 Start screen p 14 From the start screen it is not possi ble to get back to the las...

Page 17: ...ving it out of the blue colored area UV VIS II Open Mode 28 10 2014 10 57 Fig 12 Editing the favorites bar 5 3 4 Option buttons and check boxes Option buttons are framed in black Activated option butt...

Page 18: ...ple Place Printer Sound Memory Settings Date Time Region Language Select region AR Argentina AT sterreich AU Australia BE Belgique BE Belgi BG Fig 13 Predefined list UV VIS II Open Mode 28 10 201 0 0...

Page 19: ...ation of an absorbance with a user defined factor After having selected this method a window will appear asking for the factor and the wavelength for the measurement and calculation Confirming with op...

Page 20: ...rming the entries by pressing will open the measurement window Inserting the zero solution and pressing will initiate the zero measurement Then the standard solution needs to be inserted Pressing will...

Page 21: ...ement of the zero solution the sample solution need to be inserted Pressing will start the measurement process After the measurement has been completed the measurement result is displayed in the resul...

Page 22: ...zero solution and pressing initiates the zero measurement After the successful measurement of the zero solution the sample solution needs to be inserted Pressing starts the measurement process After...

Page 23: ...d the icons below the result window different sample information can be added Removal of the cuvette or leaving the menu via the remainder of the task bar icons quits the measurement process and store...

Page 24: ...est for a zero solution will appear After insertion of the rectangle cuvette the measurement is initiated by pressing After the successful measurement of the zero solution the sample solution needs to...

Page 25: ...nt process A graphic depicting the wavelengths will appear The scan of the whole wavelength range can be observed live After the measurement has been completed the complete scan will be displayed in t...

Page 26: ...or type to be measured the field light source and observer are automatically filled in These can always be manually changed In this case however the selection of the color number automatically indicat...

Page 27: ...ber can be inserted via the numeric keypad for example insert 044 for the NANOCOLOR Hardness Ca Mg test If the number is correct information about the test will appear on the display Confirming with w...

Page 28: ...ed on the factory in stalled 16 GB micro SDHC card Furthermore a data backup can be performed through the menu option In this case internal data are saved on the SDHC card UV VIS II Open Mode 28 10 20...

Page 29: ...urement Type Sample Number Begin Date End date Method User Location Type Author Operator Fig 22 Measurement memory Measurement results can be printed selected or exported as csv file The memory menu a...

Page 30: ...ed as IQC measurements The IQC menu can be entered by pressing Sub sequent selection of opens the IQC memory The results saved in this memory pro vide amongst others the basis for the IQC control char...

Page 31: ...SDHC card or a connected USB drive Additionally one can recover an existing backup from the SDHC card Please note that in case of recovering an existing backup the complete instrument storage will be...

Page 32: ...local and national legal regulations EU Directive 2012 19 EU MACHEREY NAGEL disposes old instruments free of charge Note Disposal using public waste disposal facilities is not permitted In case of dis...

Page 33: ...icates and Material Safety Data Sheets Customized special products Easy ordering and short delivery times Large number of catalogs flyer and product information of course free of charge Contact Techni...

Reviews: