20
Quickstart guide NANOCOLOR
®
UV
/
VIS
II V 1.02 / 08.15
6.1.2 Standard
c=E
x
c
s
E
s
c=E
x
c
s
The basic function “Standard” enables the definition of a sample’s concentration
on basis of the concentration of a standard solution and a zero solution. After
having selected this method, a window will appear asking for the standard con-
centration and the wavelength for the measurement.
Confirming the entries by pressing
will open the measurement window. Inserting the
zero solution and pressing
will initiate the zero measurement. Then, the standard
solution needs to be inserted. Pressing
will start the sample measurement.
After the standard solution the sample solution has to be inserted. Pressing
will
start the sample measurement. After the measurement has been completed, the mea-
surement result will be displayed in the result window. In case more samples need to be
investigated the next sample can be inserted. Pressing
will start the measurement
of the next sample.