![Macherey-Nagel NANOCOLOR VIS II Quick Start Manual Download Page 6](http://html2.mh-extra.com/html/macherey-nagel/nanocolor-vis-ii/nanocolor-vis-ii_quick-start-manual_3211133006.webp)
6
Quickstart guide NANOCOLOR
®
UV
/
VIS
II V 1.02 / 08.15
1.2 Technical details
NANOCOLOR
®
UV
/
VIS
II
NANOCOLOR
®
VIS II
Type:
Spectrophotometer with reference detector technology (RDT)
Light sources:
Halogen lamp (visible range)
Deuterium lamp (UV range)
Halogen lamp
Optical system:
Monochromator
Wavelength range:
190–1100 nm
320–1100 nm
Wavelength accuracy:
± 1 nm
Wavelength resolution:
0.1 nm
Wavelength calibration:
Automatic
Wavelength selection:
Automatic, barcode, manual
Scan speed:
900 nm / < 1 min, 1 complete
scan < 1 min
1 complete scan < 1 min
Spectral bandwidth:
< 2 nm
< 4 nm
Photometric range:
± 3.0 E in wavelength range
200–900 nm
± 3.0 E in wavelength range
340–900 nm
Photometric accuracy:
0.005 E bei 0.0–0.5 E; 1 % bei 0.5–2.0 E
Photometric linearity:
< 0.5 % bei 2 E; ≤ 1 % bei > 2 E
Stray light:
< 0.05 %
< 0.5 %
Measuring modes:
More than 200 preprogrammed tests and special methods,
100 user defined methods, absorbance, transmission, factor,
kinetics, 2-point calibration, scan, nephelometric turbidity
measurement
Turbidity measurement:
Nephelometric turbidity measurement, 0.1–1000 NTU
Cuvette slot:
Test tubes 16 mm OD
Rectangular cuvettes
2 mm, 10 mm, 20 mm, 40 mm and 50 mm
Data memory:
16 GB Micro SDHC card, 5000 measured data sets / spectra,
GLP conform
Display
HD 10.1 inch, anti-reflective cover glass with projected
capacitive touch screen (PCAP) and background LED lighting
Operation:
Barcode technology, icon-based menu guidance, touch screen
Languages:
DE / EN / FR / ES / PT / PL / HU
External light:
Insensitive, open cuvette slot
Interfaces:
LAN (CAT 6; only use shielded cable of max. 20 m length)
2 x USB (Host), 1 x USB (Function) and 1 x RS 232 (only use
shielded cable of max. 3 m length)
Update:
Via Internet / PC and USB stick
Operating range:
10–40 °C, max. 80 % relative humidity (without condensation)
Power supply:
110–240 V, ~ 50/60 Hz
12 V, DC 3A
Summary of Contents for NANOCOLOR VIS II
Page 1: ...Smart photometry NANOCOLOR VIS II NANOCOLOR UV VIS II Quickstart guide Distributed By...
Page 2: ......
Page 4: ......