This section describes the LSP installation and removal procedures, connections required for system
operation, and coated panel preparation.
LSP Removal and Installation
The Acuity LS device is delivered fully assembled with LSP installed on base. The following instructions
are only needed for replacement of an LSP. The replacement LSP is delivered with a new O-ring seal
installed. For set up of new Acuity LS device please skip to next section "LSP Surface Preparation for
Testing".
LSP removal:
1. Thoroughly rinse and dry the Acuity LS device to remove corrosive contaminants and salts,
especially near LSP and base interface.
2. Remove all socket-head cap screws using
9/64” hex wrench. Take care to avoid sideways motion
of the LSP as the screws are removed as this may damage the internal spring pin connector.
3. The LSP can then be removed from the base by lifting it straight up.
4. Once the LSP is removed, examine the base spring contacts for any corrosion or mechanical
damage. Lightly press on the spring pins with a clean wipe (no fingers) and verify that they
actuate smoothly and all return to the same position. The spring contacts on the base should
have no signs of corrosion or mechanical damage. Please contact Luna if there is any evidence
of seal leakage allowing contaminants into the device and connector.
LSP installation:
1. Ensure that the base and LSP mating surfaces are clean, dry, and free of any contaminants and
obstructions that may prevent them from mating together and forming a seal.