Operating manual – English
B - 15
CAL 2:
The offset for sensor 2 (relative humi-
dity) is set with CAL 2 (one-point calibration).
The offset turns the characteristic around the
lower calibration point (11 % r.H.). The calibra-
tion point must be in the range from 30 % to
95 % r.H. The offset is displayed in the upper
display. A maximum ± 10 % r.H. can be set.
CAL 2 can only be selected in connection
with the unit % r.H. and at a prevailing
humidity of at least 30 % r.H.
Selection can be made with
㒋
and
㒊
, confirmation with
㒉
.
The factory settings are obtained by setting the offset to 0.0.
%r.H.
dp°C dp°F
%r.H.
dp°C dp°F
CAL 2
TH
IR
DP
CFG
measured temperature value
displa
yed temperature v
alue
fact
ory calibration
new character
istic
11 % r.H.
measured humidity value
displa
yed humidity v
alue
fact
ory calibration
new character
istic
offset input
offset input
CAL 1
characteristic
is shifted
CAL 2
characteristic
is turned