Model 3030 & 3030E Series
Technical Manual
Section 4
Ludlum Measurements, Inc.
Page 10
November 2017
B-WN
: a multi-turn potentiometer used to vary the upper beta pulse
discriminator from
BT
setting to
AT
setting, normally set at 50 mV
B-TH
: a multi-turn potentiometer used to vary the lower beta pulse
discriminator from 2 to 15 mV, normally set at 4 mV
OL
: a multi-turn potentiometer that provides a means to vary the detector
current overload set point
HVL
: a multi-turn potentiometer used to calibrate the digital high voltage
HVM
: a multi-turn potentiometer used to manually adjust the high voltage
when digital control is not desired
Internal Controls
The following dipswitches and jumpers are located on the
amplifier/processor board and are only accessible when the cover is
removed.
Dipswitch 1
: When in the
OFF
position, the Model 3030 is locked into Scaler
mode only. The default position is
ON
.
Dipswitch 2
: When in the
ON
position, the calibration date is disabled. The
default position is
ON
.
HV Jumper
: When shunt is on the two outside pins, the HV is adjusted
manually with the HVM potentiometer. The normal shunt position is on the
two inside pins.
Summary of Contents for 3030
Page 4: ......
Page 5: ......
Page 6: ......
Page 7: ......
Page 8: ......
Page 9: ......
Page 56: ...Model3030 3030ESeries TechnicalManual Section7 Ludlum Measurements Inc Page 44 November 2017 ...
Page 57: ...Model3030 3030ESeries TechnicalManual Section7 Ludlum Measurements Inc Page 45 November 2017 ...
Page 66: ......
Page 67: ......
Page 68: ......
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......