Model 702i and 711i
Technical Manual
Section 9
Ludlum Measurements, Inc.
Page 9-16
December 2018
Set Date/Time
Because the Model 702i and 711i records the date and time of every spectrum collected
in Universal Coordinated Time (UTC), it must not only know the date and time but also
the current time zone. All of these are set from this menu item, which opens the
dialogue shown in Figure 31.
Figure 31 - Setting Date and Time
Use the L
EFT
and R
IGHT
arrows to move from one field to the next. To increase or
decrease the value of a particular field, use the U
P
or D
OWN
arrows, respectively. It is
recommended that if you need to change the time zone, you do this first. The hour (and
possibly minute) will be automatically updated when you make this change. Finally,
press E
NTER
to accept the currently shown date and time.
Network Setup
The Model 702i and 711i supports Ethernet communications for remote data collection,
as described in Section 3. In order to establish communications between the PC and the
Model 702i or 711i unit, the network information for the Model 702i or 711i must first
be configured in the device, then entered into the software’s “Hardware Search” utility.
Figure 32 shows the
Network Setup
dialogue, with sample values entered. You will need
to get the correct values for your own network from your network administrator.