D.1 Firmware Known Issues
F1.9.4268
Issue
Date fault triggered every time batteries are removed.
Description
When battery power is removed long enough for clock to reset, device will
show date fault ("dAt") even if calibration expiration DIP switch is turned
off.
Workaround
Acknowledge fault. Device will continue normal operation.
Issue
Instrument always starts in rate mode.
Description
Instrument always starts in rate mode, even if rate mode is disabled.
Workaround
Change modes once after power-on. Instrument will then follow config-
ured mode pattern.
Issue
Button and low tone frequencies cannot be changed.
Description
The frequency of the high tone, low tone, and button tone are config-
urable through software. A bug prevents writing changes to high tone
and button tone.
Workaround
None. Will be fixed in a future release.