![Ludlum Measurements 25024424 Manual Download Page 35](http://html.mh-extra.com/html/ludlum-measurements/25024424/25024424_manual_1946957035.webp)
C
H
A
P
T
E
R
6
F
AULT
D
ETECTION
6.1 Instrument Faults
This instrument has several faults it can report to the user. A fault is signified by flashing in
order of the red, orange, and green LEDs in addition to a unique fault tone. The display will
also cycle through the active faults.
If a fault cannot be acknowledged, the display will flash the fault code. If a fault can be
acknowledged, the fault code will be solid.
OK button:
Momentarily pressing the OK button will silence the tone and ac-
knowledge the currently displayed fault if applicable.
If the cause of a fault condition is removed, the instrument will automatically clear the fault
and continue normal operation.
6.1.1 Detector Loss of Counts
If the detector stops detecting radiation for a settable number of seconds, the instrument will
flash "LoC" or Loss of Counts. This indication is common if the unit is powered up without
a detector connected. If this indication is observed with a connected detector, remove the
unit from service and have it evaluated by a qualified repair and calibration technician.
Figure 6.1: Fault - Loss of Counts.
31
Summary of Contents for 25024424
Page 2: ......
Page 4: ......
Page 6: ......
Page 10: ......
Page 12: ......
Page 14: ......
Page 16: ......
Page 19: ...C H A P T E R 3 QUICK REFERENCE 3 1 Instrument Front Diagram 15 ...
Page 20: ...16 CHAPTER 3 QUICK REFERENCE 3 2 Instrument Rear and Sides Diagram Ludlum Measurements Inc ...
Page 21: ...3 3 INSTRUMENT INTERNAL DIAGRAM 17 3 3 Instrument Internal Diagram June 2022 ...
Page 22: ......
Page 26: ......
Page 34: ......
Page 38: ......
Page 58: ......
Page 64: ......
Page 70: ......
Page 72: ......
Page 76: ......
Page 88: ......
Page 90: ......
Page 91: ......