5.5. RATE MODE
25
Both the alert and the alarm are automatically re-armed when the readings fall below their
respective thresholds.
OK button: When a new alert and/or alarm is tripped, a press of the OK but-
ton will acknowledge the alert and/or alarm to silence the audio instead of the
normal mode specific function.
• An ALERT will flash the orange LED and sound a tone. An alert can be silenced.
• An ALARM will flash the red LED, sound a tone, and flash the ALARM icon. An alarm
can be silenced.
• A FORCED ALARM will activate when the reading exceeds the user configured maxi-
mum display. A forced alarm is similar to an ALARM with the addition of the maximum
value or 888 flashing on the display. A FORCED ALARM CANNOT be silenced.
Figure 5.1: Alert example in rate mode.
Figure 5.2: Alarm example in rate mode.
5.5 RATE Mode
In RATE mode, the current count rate is displayed. The rate is averaged based on the re-
sponse rate setting.
OK button:
Toggles click audio (or other activity indicating audio). If there
are alerts/alarms, the alert/alarm acknowledgement takes precedent and will
register the button press instead (see Alerts and Alarms on page 24).
June 2022
Summary of Contents for 25024424
Page 2: ......
Page 4: ......
Page 6: ......
Page 10: ......
Page 12: ......
Page 14: ......
Page 16: ......
Page 19: ...C H A P T E R 3 QUICK REFERENCE 3 1 Instrument Front Diagram 15 ...
Page 20: ...16 CHAPTER 3 QUICK REFERENCE 3 2 Instrument Rear and Sides Diagram Ludlum Measurements Inc ...
Page 21: ...3 3 INSTRUMENT INTERNAL DIAGRAM 17 3 3 Instrument Internal Diagram June 2022 ...
Page 22: ......
Page 26: ......
Page 34: ......
Page 38: ......
Page 58: ......
Page 64: ......
Page 70: ......
Page 72: ......
Page 76: ......
Page 88: ......
Page 90: ......
Page 91: ......