LSI LASTEM E-Log – User’s manual
57
6.3
Error messages
E-Log indicates errors by triggering the red led
Err
located on the instrument's keyboard: the kinds
of errors, listed in the following table, are indicated by the blinking modes.
Number of
blinkings
Type of problem
Troubleshooting
1
Data memory access
Try to transfer the data elaborated by the instrument;
then send the configuration again to the instrument
using
3DOM
(see §3.1.4); if the error occurs again,
contact LSI LASTEM technical support
2
IPC
Transfer the data elaborated by the instrument; then
switch the instrument off (before) and on (after); if the
error remains after many attempts, contact LSI
LASTEM technical support
3
Acquisition
from
sensors
Use
3DOM
to check the congruence of the acquisition
parameters
of
the
measures
in
the
current
configuration; send the configuration again to the
instrument (see §3.1.4); should a new error occur,
contact LSI LASTEM technical support for assistance
including the current configuration file
4
Parameter configuration Transfer the data elaborated by the instrument; then
send the configuration again to the instrument using
3DOM
(see §3.1.4); if the error occurs again, contact
LSI LASTEM technical support for assistance
including the current configuration file
5
Search data in memory If the error occurs immediately after switching on the
instrument, reset the error and continue using the
instrument as usual; should a new error occur while
operating the instrument with no communication with
the PC, contact LSI LASTEM technical support;
should the error occur while requesting the elaboration
in memory by the program
3DOM
, try requesting the
data again using a different date/time (try using a
previous date first; should a new error occur, try later
dates); should the error persist, switch the instrument
on and off and request all data indicating a date/time
definitely prior to the first date in memory; upon
completion of the data transfer, cancel them; should the
error occur again send the configuration of the
instrument to LSI LASTEM technical support
describing the problem.
6
Protocol CISS
Use
3DOM
to check the congruence of contents
parameters in the current configuration; send the
configuration again to the instrument (see §3.1.4);
should a new error occur, contact LSI LASTEM
technical support for assistance including the current
configuration file
When the instrument finds an error, it also indicates it as follows:
Summary of Contents for E-Log
Page 1: ...Cod INSTUM_00351_en E Log User s manual Updated on 09 18 2017...
Page 15: ...LSI LASTEM E Log User s manual 15...
Page 60: ...LSI LASTEM E Log User s manual 60 6 6 Mask of terminal board...
Page 61: ...LSI LASTEM E Log User s manual 61 6 7 Connection cables...
Page 62: ...LSI LASTEM E Log User s manual 62 6 8 Declaration of Conformity...