LSI LASTEM E-Log – User’s manual 

 

34 

5.4.

 

Storing elaborated data 

 
E-Log stores the calculated statistical processings (elaborated data) in the 2-MB internal memory; a 
part of this memory (128 kB) is used for configuration information and other internal information; 
the real capacity is therefore slightly less than the total storage capacity. 
 
Data storage takes place circularly; once the memory is full, the new data will replace old ones. 
 
By  sending  a  new  configuration  to  the  instrument,  all  data  so  far  stored  will  be  cancelled;  this 
happens because the new configuration information might be potentially not in tune with them, and 
the PC might misinterpret them. 
 
The instrument stores the data in the internal memory only when a 

data page

 is full; the dimension 

of  this  page  equals  to  256  bytes;  therefore  the  instrument  might  store  the  data  only  after  several 
sequences of elaboration; this depends on the programmed elaboration rate, on the number of active 
measures  and,  for  each  one  of  them,  on  the  selected  elaboration  elements.  Please  note  that  by 
switching  off  the  instrument,  some  elaborated  data  inside  the 

data 

page  and  those  not  definitely 

stored might be lost; for this reason we advise transferring the elaborations which have not yet been 
transmitted to the PC before switching off E-Log. 
 
 

5.4.1

 

Memory autonomy 

 
Depending  on  the  chosen  configuration  (measures,  types  of  elaboration  for  each  measure  and 
acquisition rate) the instrument will work somewhat autonomously as far as the maximum storage 
time is concerned, without replacing old data with new ones. The calculation of the time dimension 
of the maximum number of storable data is as follows: 
 
A = K / (86400 / RE * NE) 
 
where: 
 
A =  number of days of autonomy of the data memory; 
K =  value subordinate on the size of the used memory; for the 2 MB internal memory, K=430592 

(from  2.12.00  instrument  firmware  version,  K=415744);  for  the  8  MB  internal  memory, 
K=1841152; 

RE = elaboration rate expressed in seconds; 
NE = total number of elaboration’s elements programmed for all measures.  
 
 
 

5.5.

 

Actuation logics 

 
E-Log has got an actuation logics library. It’s useful to switch-on all type devices (alarms, solenoid 
valves,  motors)  according  the  surveyed  parameters  in  surrounding  environment.  The  actuation 
logics are based on the instantaneous  value of the acquired and calculated measures. They  can be 
programmed  up  to  20  calculation  algorithms,  that  use  same  or  different  logics.  One  or  more 
algorithms can be combined in two differed modes , in order to switch-on the selected actuator: 
 

Summary of Contents for E-Log

Page 1: ...Cod INSTUM_00351_en E Log User s manual Updated on 09 18 2017...

Page 2: ...this manual may be changed without prior notification No part of this manual may be reproduced neither electronically or mechanically under any circumstance without the prior written permission of LSI...

Page 3: ...g additional features 26 5 1 Starting and running the survey 26 5 1 1 Checking the power supply voltage 26 5 2 Acquisition and calculation of measures 26 5 2 1 Acquisition from sensors with serial or...

Page 4: ...ransmission using TCP IP 47 5 6 10 E Log connected like master slave mode 47 5 6 11 E Log with ZigBee radio 48 5 7 Operating at low energetic consumption 51 6 Appendixes 53 6 1 Technical specification...

Page 5: ...iner Power the instrument in a suitable manner Connect the instrument to the power supply indicated in the model in your possession Carry out all connections in a suitable manner Pay strict attention...

Page 6: ...nvironmental protection and collection LSI LASTEM advises to handle E Log as waste of electrical and electronic equipment WEEE It is therefore not to be collected with any other kind of waste LSI LAST...

Page 7: ...and the terminal connections to power the instrument power supply or battery The terminal board is protected by a carter Picture 1 that covers the terminals remove it by pressing slightly on both sid...

Page 8: ...he right hand side panel in this case the positive pole is the one inside the connector In any case pay attention not to invert the power polarity although the instrument is protected from wrong proce...

Page 9: ...e to consider both the energetic saving and the time the sensor needs to initialize The association between input and switching power is fixed as shown in the table below The number of the terminal is...

Page 10: ...o return a voltage signal from the current generated by the sensor Program 3DOM supplies a library for the setting of the LSI LASTEM sensors including some powered outputs for such models the settings...

Page 11: ...ized output Picture 6 Moreover if you want connect to E Log some LSI LASTEM portable probes type BST BSU BSO or sensors not branded LSI LASTEM which mount minidin connectors you have to use ELA115 int...

Page 12: ...All sensors generating tension or current signals with or without external power can be connected to single ended inputs see connection scheme Sensors with analogue signal single ended mode The drop...

Page 13: ...ng to E Log model Both serial ports leave our factory with the following default configuration Baud rate 9600 bps Data bit 8 Stop bit 1 Parity none Network address 1 Flow control RTS signal only The i...

Page 14: ...med with 3DOM 3 2 Configuring the operative modes LSI LASTEM supplies the instrument with a standard configuration In order to modify the configuration parameters to one s specific needs connect a PC...

Page 15: ...LSI LASTEM E Log User s manual 15...

Page 16: ...ouses the keyboard bright indicators led and the display used to check if the instrument is working properly The lower part includes a carter to protect the electrical connections to the terminals loc...

Page 17: ...easure list Holds releases measure scrolling Switches the display off and on see 4 3 4 While showing diagnostic information Shifts to the display window of the values of the measures If applicable it...

Page 18: ...uuuuuuuu serial number or instrument number settled by the user These informations are also available while the instrument is working and can be called by simply using the keyboard to select this win...

Page 19: ...ow of the system in order to decode this error go to 6 3 mmmm data memory capacity in kByte wwww hexadecimal number that expresses the value of the position writing in the elaborated data memory the s...

Page 20: ...network address of the instrument yyyyy communication rate bit rate of the serial port The character shows the current serial port for this port are valid the modification commands of the bit rate fr...

Page 21: ...ice if the forcing is also imposed to an actuator state that can assume 0 or 1 value means that actuator will never be change by any internal logic except the commands received from remote through CIS...

Page 22: ...ers its use only for serial port 1 alternatively the display goes to the next diagnostic window In case the configuration considers the use of the GPRS modem but the modem is switched off the window s...

Page 23: ...o No Conn status of not established connection to the PAN o Fail radio module absent or not working x indicates the modem status it can take these values o 0 the modem has undergone a hardware reset...

Page 24: ...orrect transmission o 2 failed transmission o 22 address node not valid o 33 failed confirmation of network acknowledge o 34 no network connection o 35 addressing of network node obtained indipendentl...

Page 25: ...cation indicator Description Off Instrument switched off or low feeding voltage min 7 V Single quick green blinking Instrument switched on Regular operation acquisition mode Three quick green blinking...

Page 26: ...it must be 7 V at least In case it isn t enough E Log closes the survey in progress and stands by During the stand by it checks every minute the power supply voltage until it ll reach the minimum val...

Page 27: ...Gill protocol Gill format polar continuous E Log supports the connection to one sensor 3DOM software programs the used protocol it isn t possible use more than one protocol and for this reason it isn...

Page 28: ...nsor equivalent to character ASCII Q is considered by E Log like value 1 the next letter R like value 2 and so on 3DOM application automates some configuration operations by means of its sensors libra...

Page 29: ...d direction and relative humidity are excluded Linearized quantity control the instrument produces null output when receiving null input All above parameters are indicated in section Measure propertie...

Page 30: ...ndicating among others its serial number When operating in this acquisition mode the instrument s LED will blink differently see 4 4 Please note that in this condition the instrument Acquires all sens...

Page 31: ...of all primary measures that allow the estimate of calculated measures a calculated measure can be a primary measure for new calculated measure 2 Collection of the value of primary measures if at lea...

Page 32: ...percentage It is not possible to combine arithmetical and vectorial calculations for each specific measure Programm 3DOM can be used to program the elaboration parameters The instrument doens t have i...

Page 33: ...gra Resulting average direction Dmr Vel Dir rad Cos Vel Dir rad Sin gra atan2 Resulting average speed Vmr n Vel Dir rad Cos Vel Dir rad Sin 2 2 Direction s standard deviation sigma theta DevStandDir 2...

Page 34: ...ht be lost for this reason we advise transferring the elaborations which have not yet been transmitted to the PC before switching off E Log 5 4 1 Memory autonomy Depending on the chosen configuration...

Page 35: ...uator even if this occurs temporally after the deactivation The programming of the actuation logics happen by means of the 3DOM program according to two different phases 1 Selection of the logics and...

Page 36: ...he surface of the water doesn t allow the right evaporation in the morning and in the evening 5 5 3 Start precipitation alarm The logic uses a measure connected to a rain gauge to detect the start pre...

Page 37: ...s both univocally and simultaneously To the threshold values can be applied a further hysteresis value this can avoid continuous alarm state changes in case the measure value moves nearly around the t...

Page 38: ...and at the same time lesser than the maximum threshold value return to no alarm condition when the measure value is lesser than the minimum threshold value or greater than the maximum threshold hyster...

Page 39: ...ust be consecutively ordered the list cannot include any measures that shouldn t be considered in the comparing because the logic programming needs the ordinal numbers of the first and last measure Th...

Page 40: ...off times If timer type is not cyclic the day time of power on the actuator and the day time of power off This logic is joinable with other logics in AND mode for example to allow the alarms activatio...

Page 41: ...ith GSM or GPRS modem in this case follow the instruction in 5 7 to reduce the modem power consumption To execute local communications with E Log remove the connection between the modem and serial por...

Page 42: ...ine sensor Boschung sensor data acquisition from Boschung sensor Some LSI LASTEM applications like i e InfoPanel are able to obtain the instantaneous measures data from serial port 2 In this way the s...

Page 43: ...ral Acquisition of sampled instantaneous values and transmission from connected instrument X Komoline Acquisition of sampled instantaneous values and transmission from connected instrument X Boschung...

Page 44: ...ses the GPRS data packet transmission and TCP or FTP protocols DEA717 DEA718 with E Log FW until V 2 29 00 DEA718 1 with E Log FW V 2 30 00 or later to allow the continuous transmission with transmiss...

Page 45: ...eceived data of some parameters you must follow the following tables OVM Obstruction to Vision Message Value Meaning 0 No obstruction 1 HZ Haze 2 FG Fog SFM Self test and Monitoring Subparameter Forma...

Page 46: ...two different transmission s modes Through TCP socket Through FTP protocol The TCP socket connection allows to have greater control over the datalogger against the FTP protocol because not only the el...

Page 47: ...nding and closing file Check the log recorded by the FTP server 28 30 33 GPRS context creation and network access Using 3DOM verify the APN connection parameters against the SIM used inside the modem...

Page 48: ...ters of port 2 Network Address 1 and Spontaneous transmission Rate 0 input one rate lower than the acquisition rate of fast serial measure set up on master we recommend the input of slave s transmissi...

Page 49: ...plications buildings and plants for outdoor applications that the signal encounters during their journey to the destination of communication Specifications Performances Indoor standard range up to 60...

Page 50: ...install the network devices Put the devices in place making attention to distance as the crow files among respective nodes remain within max range specified in specifications and considering the obsta...

Page 51: ...structure through Master any time you need it Note 1 in a network in case of reprogramming of Master without modify of configuration of Slave instruments which are turned on and active in acquisition...

Page 52: ...so they aren t operating Disable communication protocol on serial port 2 in case it isn t used or in case it corresponds to following types TTY Modbus Gill anemometer ATTENTION Default configuration...

Page 53: ...er Temperature drift scale 300 1200 mV 0 01 FSR 10 30 C scale 39 mV 0 01 FSR 10 30 C scale 78 mV 0 01 FSR 10 30 C Digital inputs Number of digital inputs 4 like follow 2 inputs for sensors with optoel...

Page 54: ...Power Input voltage DC 10 14 V Average power consumption without probes feed 12 V 8 channels acquisition display power on 136 mW 8 channels acquisition display power off 115 mW Power down 4 mW Protec...

Page 55: ...ximum o Totalization o Angular average Thermo hygrometric quantities UNI EN ISO 7726 ISO WD 7730 VDI 3786 o Relative humidity with psychrometric calculation dry humid bulb o Absolute humidity o Specif...

Page 56: ...ociated with 3DOM software from 3 8 version These measure are set a fixed update time of 1 second the update of value of calculated measure on the actuator is always 1 second delay compared to the sta...

Page 57: ...ument using 3DOM see 3 1 4 if the error occurs again contact LSI LASTEM technical support for assistance including the current configuration file 5 Search data in memory If the error occurs immediatel...

Page 58: ...ers that the error has already been found by the person using the computer In both cases the reset operation will locally switch off the optical indicator Err eliminate the error number from the diagn...

Page 59: ...ation with PC With or without terminals for connection of sensor cables fixed or removable With or without alphanumerical display and keyboard Code Power supply 12 Vdc Built in Zigbee radio Fixed term...

Page 60: ...LSI LASTEM E Log User s manual 60 6 6 Mask of terminal board...

Page 61: ...LSI LASTEM E Log User s manual 61 6 7 Connection cables...

Page 62: ...LSI LASTEM E Log User s manual 62 6 8 Declaration of Conformity...