7 Alarms and faults
7.5 Technical alarms
DISPLAY
CAUSE
ACTION
Service necessary.
Please get in touch
with your specialist
dealer/contact.
Technical fault which can only
be eliminated by an
authorized specialist dealer.
Have device repaired.
Battery defective.
Service necessary.
Battery defective.
Have battery replaced.
Device defective.
Have device repaired.
Battery not present.
Service necessary.
Battery defective.
Have device repaired.
Unapproved battery in use.
Battery capacity highly
critical
Battery discharged (less than
5 % capacity remaining).
Connect the device to the power
supply.
Battery capacity
critical
Battery discharged (less than
10 % capacity remaining).
Connect the device to the power
supply.
Battery switched off
due to temperature
Battery too hot.
Operate device at an ambient
temperature of 5 °C to 40 °C.
Service life of battery
ended. Have battery
replaced.
Service life of battery ended. Have battery replaced.
Battery temperature
high
Battery too hot.
Operate device at an ambient
temperature of 5 °C to 40 °C.
Battery not detected.
Service necessary
Battery defective.
Have battery replaced.
Device defective.
Have device repaired.
Intake area covered.
Please keep intake
area free.
Intake area covered.
Keep intake area free.