8 Servicing
8 Servicing
8.1 Safety information
8.2 General information
• Only service the device when no patient is connected to it.
• Have measures such as repairs, maintenance, and servicing, as well as modifications to
the device, carried out exclusively by the manufacturer or by specialists expressly
authorized by the manufacturer.
• The device is designed for a service life of 6 years. If used in accordance with the
intended use, the device requires no servicing during this period. If the device is used
beyond this period, it needs checking by an authorized specialist dealer.
• For Germany: In accordance with §11 of the German law governing the owners/
operators of medical devices, the device must be subjected to a Technical Safety Check
[Sicherheitstechnische Kontrolle (STK)] every 2 years. Country-specific requirements
apply to all other countries.
• If the device has a battery, this must be replaced every 4 years.
9 Transport and storage
Store and transport the device under the specified ambient conditions. Clean the
device before storing it.
If the device has an internal battery that is always supposed to be ready for use, leave
the device connected to the power supply. This ensures that the battery is always fully
charged.
If the device is not connected to the power supply for an extended period, the battery
will discharge. We recommend checking charge status regularly and recharging the
battery with the aid of the device (if required).
Risk of injury from modified ME equipment!
An unauthorized modification to the ME equipment may put the patient at risk.
Do not modify the device without the manufacturer’s consent.
If the device is modified, carry out the relevant investigations and tests