5 Settings in the menus
5.2.4 Patient menu - softSTART/softSTOP
To call up the softSTART/softSTOP menu, the device must be on standby. If enabled by
the physician or specialist dealer, the following parameters can be set here:
PARAMETER
ADJUSTABLE
VALUES
DESCRIPTION
softSTART time
T
5-minute increments
within the framework
specified by the
physician or specialist
dealer (e.g. 5 min. to a
maximum of 45 min.).
Here you can set the time for which
ventilation pressure rises to therapy
pressure during softSTART.
If this function cannot be selected, it
needs to be enabled by your physician or
specialist dealer.
softSTART EPAP
pressure
0.2 hPa increments
within the framework
specified by the
physician or specialist
dealer (e.g. at least
4 hPa to 25 hPa).
Here you can set the pressure on
exhalation with which softSTART will
begin.
If this function cannot be selected, it
needs to be enabled by your physician or
specialist dealer.
softSTOP time
T
5-minute increments
within the framework
specified by the
physician or specialist
dealer (e.g. 5 min. to a
maximum of 45 min.).
Here you can set the time period during
which ventilation pressure is reduced
within the softSTOP framework.
If this function cannot be selected, it
needs to be enabled by your physician or
specialist dealer.