34
Each digit indicates the following status:
D15
Start Word = 02
D14
4
D13
Sending upper display data = 1
Sending lower display data = 2
D12, D11
Annunciator for Display
µS = 13
mS = 14
ppm = 19
pH = 05
mV = 18
mg/L = 07
% O2 = 06
D10
Polarity
0 = Positive 1 = Negative
D9
Decimal Point (DP) position from right to left
0 = No DP, 1= 1 DP, 2 = 2 DP, 3 = 3 DP
D8 to D1
Display reading, D1 = LSD, D8 = MSD
For example:
If the display reading is 1234, then D8 to
D1 is: 00001234
D0
End Word = 0D
RS232 setting
Baud rate
9600
Parity
No parity
Data bit no.
8 Data bits
Stop bit
1 Stop bit