Manual range operation
The meter default is set to auto range mode. Under auto range
measurement, the bottom right display will show the “Auto
Range“ indicator. If manual range mode is required, the procedures
are as follows:
1) Press the “Range” button continuously for at least two seconds
until the bottom right display shows the “Manual Range“ indica-
tor. Release the “Range” button. Now the meter is set for manual
range operation.
2) Press the “Range” button once to change the range. The range
(200 µS, 2 mS, 20 mS, 200 mS) will show under the measure-
ment value.
3) • If the display shows “
----
“, it indicates an overrange.
Select the next, higher range.
• If the display shows “ ---- “, it indicates an underrange.
Select the next, lower range.
4) To change Manual Range back to Auto Range, press the
“Range” button continuously for at least two seconds until the
bottom right display shows the “Auto Range“ Indicator. Release
the “Range” button. Now the meter is set for Auto range mode
again.
Change the Temp. unit to °F
To change the Temp. unit from °C to °F, please refer to page 28,
chapter 7.6 (Temp. Unit Default Setting)
Change the Temp. Coefficient Factor
The default Temp. compensation factor value of the measurement
solution is set to 2.0% per °C. To change it, please refer to page
28, chapter 7.8 (Temp. Compensation Factor Setting).