background image

Lucetta™ 2 Luminometer Manual – 

Bioscience Solutions

   9

4.0 Lucetta™ 2 Luminometer  
Operation

4.1  Menu

On the front face of the Lucetta™ 2 Luminometer is an LCD display 
showing the menus, the measurement parameters, and the 
measurement results. To open a submenu from the main menu (Figure 
4) move the gray select bar up and down by using the two arrow 
buttons 



. Once at the desired position, press OK to activate that 

choice.

Figure 4. Main menu of the Lucetta™ 2 Luminometer. 

Run MycoAlert PLUS

” or “

Run MycoAlert”

 will take you to the programs 

for the MycoAlert™ PLUS or MycoAlert™ Mycoplasma Detection Assay, 
respectively. From here, the MycoAlert™ PLUS/ MycoAlert™ Assay can 
be performed and the contamination status of cells be determined. 

Run Single Read

” will take you to the single read programs. These 

6 customizable programs can be used to perform the ViaLight™ Cell 
Proliferation and Cytotoxicity Assay or the ToxiLight™ Cytotoxicity 
Assay.

Results

” will allow you to view stored data and upload the data from 

selected runs to your PC via the integral USB port on the rear side.

Settings

” opens a submenu that allows you to change date/time, 

review the instrument information and adapt Single Read parameters. 
End user settings can be protected with a password here. The submenu 
“MycoAlert Settings” is only accessible by Lonza personnel.

4.2  MycoAlert™ PLUS/ MycoAlert™ Run

Select the correct program for the assay you want to run as 
interpretation of the results is slightly different. For a detailed 
description of the assay steps, please refer to the MycoAlert™ PLUS/ 
MycoAlert™ Assay Instructions.

1.  Turn on the instrument by pressing the ON button.
2.  Use arrow buttons 



 to select RUN MYCOALERT PLUS when 

running a MycoAlert™ PLUS Mycoplasma Detection Assay or 
RUN MYCOALERT when running a MycoAlert™ PLUS Mycoplasma 
Detection Assay, respectively. Press OK.

Note

Pressing ESC in any of the process related screens will exit the 
measurement process and bring you back to the start screen. Do not 
press ESC unless you really want to exit the measurement cycle.

3.  Enter the number of samples (1–99) by using arrow buttons 



 to select the digit you want to change. Use arrow buttons 



 to select a number from 0 to 9. Press OK. The instrument 

will first perform reading A for all samples and then continue with 
reading B for all samples. For convenient handling of incubation 
times we recommend a maximum number of five samples in one 
run. 

4.  Follow instructions to add MycoAlert™ PLUS or MycoAlert™ 

Reagent to the samples. Press OK when completed. This will start 
the automatic countdown for the required incubation time of  
5 minutes.

5.  Wait for the instrument to beep and then press OK to continue.

Note

You can abort the automatic countdown by pressing arrow button 

. You are prompted to confirm your decision by pressing the arrow 

button 

 and OK. This will access the next screen and allow you to 

time the incubation manually.

6.  Open the black lid of the Lucetta™ 2 Luminometer and insert 

the first sample vial. Upon closing the lid the measurement 
automatically starts with reading A. The measured value in RLU 
will appear on the display.

7.  Press OK to continue with the next sample.
8.  After reading A is finished for all samples press OK to continue. 

Add the MycoAlert™ Substrate to all samples and press OK when 
completed. Again, this will start the automatic countdown for the 
required incubation time of 10 minutes.

9.  Wait for the instrument to beep and then press OK to continue.

Note

You can abort the automatic countdown by pressing arrow button 

. You are prompted to confirm your decision by pressing the arrow 

button 

 and OK. This will access the next screen and allow you to 

time the incubation manually.

10. Open the lid of the Lucetta™ 2 Luminometer and insert the first 

sample vial. Upon closing the lid, the measurement automatically 
begins with reading B. The measured value in RLU will appear on 
the display.

11. Press OK to continue with the next sample.
12. After reading B is finished for all samples, press OK. The 

complete results for all samples will be displayed (starting with 
sample no.1) including the individual RLU readings A and B, the 
calculated B/A ratio and the contamination status (Figure 5).

13. Pressing the arrow buttons 



 allows you to switch back and 

forth between the sample results.

14. Results are classified as CLEAN, BORDERLINE or CONTAMINATED 

according to the values shown below. Classification is 
indicated by 3 LEDs: green = clean, yellow = borderline, red = 
contaminated.

MAIN MENU

Run MycoAlert PLUS

Run MycoAlert

Run Single Read 

Results

Settings

Summary of Contents for Lucetta 2

Page 1: ...Bioscience Solutions Lucetta 2 Luminometer Manual...

Page 2: ......

Page 3: ...ctions 5 2 2 Maintenance 5 2 3 Safety Instructions and Use Precautions 6 2 4 Storage Conditions 6 2 5 Batteries 6 2 6 Return Shipment 7 2 7 Instrument Disposal 7 3 0 Setup Instructions 7 3 1 Lucetta 2...

Page 4: ...ta Storage and PC Upload 11 5 1 Maximum Storage Capacity 11 5 2 Viewing Logged Measurements 11 5 3 Upload to a PC 11 6 0 Setting Instrument Parameters 12 6 1 Password 12 6 2 Customer Settings 12 6 3 D...

Page 5: ...abled by the Lucetta Connect software Broad range of applications Lucetta 2 Luminometer applications include MycoAlert PLUS and MycoAlert Mycoplasma Detection Assays ATP assays e g ViaLight Cell Proli...

Page 6: ...strument is not approved for use in fire or explosion potential areas nor for use with flammable or explosive media The sample holder must remain free of dust and contamination The holder should only...

Page 7: ...tance if required 2 7 Instrument Disposal Decontaminate the instrument before disposal This luminometer contains electronic parts To prevent environmental pollution please dispose of the instrument an...

Page 8: ...nd the age of the battery The battery may auto discharge when the instrument is not used for an extended period To optimize the battery lifespan the display light switches off after 15 sec if the inst...

Page 9: ...from 0 to 9 Press OK The instrument will first perform reading A for all samples and then continue with reading B for all samples For convenient handling of incubation times we recommend a maximum num...

Page 10: ...efine a new protocol 4 Each measurement is given an identifier number displayed under TEST By default the number begins with 1 counting upwards The number can be customized by using the arrow buttons...

Page 11: ...ns to select day month or year Depending on the selected country for the date format for more details see section 6 0 the format is MM DD YYYY USA or DD MM YYYY Europe Theflashingcursorindicatesthenum...

Page 12: ...word for a new instrument is 0000 Maximum number of digits for a password is four 1 Navigate down to SETTINGS and press OK 2 Select PASSWORD and press OK 3 Enter OLD PASSWORD To change the given numbe...

Page 13: ...Figure 13 Display for setting date and time 6 4 System Display brightness and contrast are adaptable to the lighting conditions 1 Navigate down to SETTINGS and press OK 2 Select SYSTEM and press OK 3...

Page 14: ...liance with operating conditions Background or measurement value too low 0 RLU s Detector or electronics damaged or incorrect instrument settings Service required please contact Lonza Unexpected or wr...

Page 15: ...asuring protocols Data storage Storage of the last 2000 results Hardware and Software Requirements for Data Upload Optional PC with industrial standard Microsoft Windows 7 32 64 bit or higher Microsof...

Page 16: ...22 75323 Bad Wildbach Germany and distributed by Lonza Cologne GmbH Nattermannallee 1 50829 Cologne Germany and Lonza Walkersville Inc 8830 Biggs Ford Road Walkersville MD 21793 Microsoft Windows and...

Reviews: