6
Bioscience Solutions
– Lucetta™ 2 Luminometer Manual
Rechargeable batteries
Nickel metal hydride (NiMH) batteries are used to wirelessly operate
the Lucetta™ 2 Luminometer.
Batteries should be allowed to fully deplete, then fully recharged every
couple of months.
Discharging duration: approx. 6 h
Recharging duration: approx. 4–5 h
2.3 Safety Instructions and Use Precautions
Lucetta™ 2 Luminometer instruments have been tested by the
manufacturer, certified by international safety standards and are safe
to use when operated in accordance with this manual.
Please adhere to the following safety instructions and use precautions
before and during operation of the system or taking the instrument
into service:
– Read and understand the the Lucetta™ 2 Luminometer Manual
prior to using the instrument. All potential users should have
access to the manual and should read and understand it prior to
using the instrument.
– The Lucetta™ 2 Luminometer can be operated using an electrical
outlet or as a portable, battery-operated instrument. Ideally,
battery charging occurs while the instrument is not in use,
e.g overnight. To charge the battery, the instrument should be
attached to the provided AC power supply/charger and turned on.
– The power supply must be connected to a wall outlet complying
with local regulations of the country of its installation and
providing voltage and current according to the specification of the
power supply.
– Do not connect the power cord near fluids to avoid electric shock
and burning. The power cord must never become wet!
–
When charging the instrument or operating with an external
power supply, make sure to use ONLY the originally provided
power supply. Substitution may create a hazard and harm the
instrument.
– Unauthorized opening, modification, repair, or improper usage of
the instrument may result in damage and will void all warranties.
– The instrument may only be used for its intended purpose (see
section 2.1).
It is the operator’s responsibility to adhere to local regulations when
installing and operating the system.
– The user must ensure the instrument is configured and installed
such that functionality is not impaired. Please refer to the setup
instructions (section 3.2).
– The instrument should be ultilized only when located on a stable
table or bench.
– Do not expose the instrument to a humid environment (>70%).
– The instrument should not be exposed to direct sunlight nor be
located in a hot environment (above 30°C).
– The instrument is not approved for use in fire or explosion
potential areas, nor for use with flammable or explosive media.
– The sample holder must remain free of dust and contamination.
The holder should only be open during sample loading/unloading
and cleaning.
– Use only specified sample tubes.
– Assays must be validated with the instrument prior to routine use.
– Certain biological and chemical components and samples may
pose a hazard or infection risk. Always refer to the description of
the sample to ensure adequate safety precautions are in place.
Utilize appropriate protective equipment to avoid exposure.
– Servicing and repair may only be performed by Lonza authorized
personnel.
– The user may perform only those maintenance steps specifically
described in this manual (section 2.2).
– The instrument must be decontaminated before repair work or
service.
– Do not enter or place foreign objects in the sample area.
– Avoid liquid spills on the instrument housing and in the sample
area. Wipe up all spills immediately and decontaminate the
instrument surfaces appropriately.
Note
If liquid gets inside the instrument, switch off the instrument
immediately. Do not operate the instrument if internal components
have been exposed to fluids. The instrument must be cleaned by Lonza
authorized personnel.
– Handling of instrument components with potential sample
contamination shall always be performed with protective gloves.
Disposal of parts must be according to federal, state or local
procedures for biohazardous waste handling and disposal. Use
secure leak proof containers and avoid unprotected handling of
parts.
Lonza Cologne GmbH and its distributors disclaim all warranties, either
expressed or implied, and shall in no event be liable for damages,
including those to third parties, caused by or arising from improper
installation or any operation or use in violation of the above safety and
handling instructions.
2.4 Storage Conditions
For shipping and long-term storage, the instrument should remain in
the original cardboard box in a dry, dust-free environment and protected
from direct sunlight and significant temperature fluctuations. Storage
conditions should remain at 5–40°C and up to 80% humidity at 30°C,
non-condensing.
2.5 Batteries
To charge the battery, the instrument is attached to the provided AC
power supply/charger and switched on. Battery charging should be
done when instrument is not in use, e.g overnight. The instrument
holds three NiMH batteries each 1.2V/4000 mAh.
In case of battery defect, please contact Lonza Scientific Support Team.
Lonza Cologne GmbH accepts no liability for damage caused by use of
other batteries than recommended by Lonza Scientific Support Team.