Bioscience Solutions

 – Lucetta™ 2 Luminometer Manual

Rechargeable batteries

Nickel metal hydride (NiMH) batteries are used to wirelessly operate 
the Lucetta™ 2 Luminometer. 
Batteries should be allowed to fully deplete, then fully recharged every 
couple of months.

Discharging duration: approx. 6 h
Recharging duration: approx. 4–5 h

2.3 Safety Instructions and Use Precautions

Lucetta™ 2 Luminometer instruments have been tested by the 
manufacturer, certified by international safety standards and are safe 
to use when operated in accordance with this manual.

Please adhere to the following safety instructions and use precautions 
before and during operation of the system or taking the instrument 
into service:

 

– Read and understand the the Lucetta™ 2 Luminometer Manual 

prior to using the instrument. All potential users should have 
access to the manual and should read and understand it prior to 
using the instrument. 

 

– The Lucetta™ 2 Luminometer can be operated using an electrical 

outlet or as a portable, battery-operated instrument. Ideally, 
battery charging occurs while the instrument is not in use, 
e.g overnight. To charge the battery, the instrument should be 
attached to the provided AC power supply/charger and turned on.

 

– The power supply must be connected to a wall outlet complying 

with local regulations of the country of its installation and 
providing voltage and current according to the specification of the 
power supply.

 

– Do not connect the power cord near fluids to avoid electric shock 

and burning. The power cord must never become wet!

 

When charging the instrument or operating with an external 
power supply, make sure to use ONLY the originally provided 
power supply. Substitution may create a hazard and harm the 
instrument.

 

– Unauthorized opening, modification, repair, or improper usage of 

the instrument may result in damage and will void all warranties.

 

– The instrument may only be used for its intended purpose (see 

section 2.1).

It is the operator’s responsibility to adhere to local regulations when 
installing and operating the system.

 

– The user must ensure the instrument is configured and installed 

such that functionality is not impaired. Please refer to the setup 
instructions (section 3.2).

 

– The instrument should be ultilized only when located on a stable 

table or bench.

 

– Do not expose the instrument to a humid environment (>70%).

 

– The instrument should not be exposed to direct sunlight nor be 

located in a hot environment (above 30°C).

 

– The instrument is not approved for use in fire or explosion 

potential areas, nor for use with flammable or explosive media.

 

– The sample holder must remain free of dust and contamination. 

The holder should only be open during sample loading/unloading 
and cleaning.

 

– Use only specified sample tubes.

 

– Assays must be validated with the instrument prior to routine use.

 

– Certain biological and chemical components and samples may 

pose a hazard or infection risk. Always refer to the description of 
the sample to ensure adequate safety precautions are in place. 
Utilize appropriate protective equipment to avoid exposure.

 

– Servicing and repair may only be performed by Lonza authorized 

personnel.

 

– The user may perform only those maintenance steps specifically 

described in this manual (section 2.2).

 

– The instrument must be decontaminated before repair work or 

service.

 

– Do not enter or place foreign objects in the sample area.

 

– Avoid liquid spills on the instrument housing and in the sample 

area. Wipe up all spills immediately and decontaminate the 
instrument surfaces appropriately.

Note 

If liquid gets inside the instrument, switch off the instrument 
immediately. Do not operate the instrument if internal components 
have been exposed to fluids. The instrument must be cleaned by Lonza 
authorized personnel.

 

– Handling of instrument components with potential sample 

contamination shall always be performed with protective gloves. 
Disposal of parts must be according to federal, state or local 
procedures for biohazardous waste handling and disposal. Use 
secure leak proof containers and avoid unprotected handling of 
parts.

Lonza Cologne GmbH and its distributors disclaim all warranties, either 
expressed or implied, and shall in no event be liable for damages, 
including those to third parties, caused by or arising from improper 
installation or any operation or use in violation of the above safety and 
handling instructions.

2.4  Storage Conditions

For shipping and long-term storage, the instrument should remain in 
the original cardboard box in a dry, dust-free environment and protected 
from direct sunlight and significant temperature fluctuations. Storage 
conditions should remain at 5–40°C and up to 80% humidity at 30°C, 
non-condensing.

2.5 Batteries

To charge the battery, the instrument is attached to the provided AC 
power supply/charger and switched on. Battery charging should be 
done when instrument is not in use, e.g overnight. The instrument 
holds three NiMH batteries each 1.2V/4000 mAh.

In case of battery defect, please contact Lonza Scientific Support Team. 
Lonza Cologne GmbH accepts no liability for damage caused by use of 
other batteries than recommended by Lonza Scientific Support Team.

Summary of Contents for Lucetta 2

Page 1: ...Bioscience Solutions Lucetta 2 Luminometer Manual...

Page 2: ......

Page 3: ...ctions 5 2 2 Maintenance 5 2 3 Safety Instructions and Use Precautions 6 2 4 Storage Conditions 6 2 5 Batteries 6 2 6 Return Shipment 7 2 7 Instrument Disposal 7 3 0 Setup Instructions 7 3 1 Lucetta 2...

Page 4: ...ta Storage and PC Upload 11 5 1 Maximum Storage Capacity 11 5 2 Viewing Logged Measurements 11 5 3 Upload to a PC 11 6 0 Setting Instrument Parameters 12 6 1 Password 12 6 2 Customer Settings 12 6 3 D...

Page 5: ...abled by the Lucetta Connect software Broad range of applications Lucetta 2 Luminometer applications include MycoAlert PLUS and MycoAlert Mycoplasma Detection Assays ATP assays e g ViaLight Cell Proli...

Page 6: ...strument is not approved for use in fire or explosion potential areas nor for use with flammable or explosive media The sample holder must remain free of dust and contamination The holder should only...

Page 7: ...tance if required 2 7 Instrument Disposal Decontaminate the instrument before disposal This luminometer contains electronic parts To prevent environmental pollution please dispose of the instrument an...

Page 8: ...nd the age of the battery The battery may auto discharge when the instrument is not used for an extended period To optimize the battery lifespan the display light switches off after 15 sec if the inst...

Page 9: ...from 0 to 9 Press OK The instrument will first perform reading A for all samples and then continue with reading B for all samples For convenient handling of incubation times we recommend a maximum num...

Page 10: ...efine a new protocol 4 Each measurement is given an identifier number displayed under TEST By default the number begins with 1 counting upwards The number can be customized by using the arrow buttons...

Page 11: ...ns to select day month or year Depending on the selected country for the date format for more details see section 6 0 the format is MM DD YYYY USA or DD MM YYYY Europe Theflashingcursorindicatesthenum...

Page 12: ...word for a new instrument is 0000 Maximum number of digits for a password is four 1 Navigate down to SETTINGS and press OK 2 Select PASSWORD and press OK 3 Enter OLD PASSWORD To change the given numbe...

Page 13: ...Figure 13 Display for setting date and time 6 4 System Display brightness and contrast are adaptable to the lighting conditions 1 Navigate down to SETTINGS and press OK 2 Select SYSTEM and press OK 3...

Page 14: ...liance with operating conditions Background or measurement value too low 0 RLU s Detector or electronics damaged or incorrect instrument settings Service required please contact Lonza Unexpected or wr...

Page 15: ...asuring protocols Data storage Storage of the last 2000 results Hardware and Software Requirements for Data Upload Optional PC with industrial standard Microsoft Windows 7 32 64 bit or higher Microsof...

Page 16: ...22 75323 Bad Wildbach Germany and distributed by Lonza Cologne GmbH Nattermannallee 1 50829 Cologne Germany and Lonza Walkersville Inc 8830 Biggs Ford Road Walkersville MD 21793 Microsoft Windows and...

Reviews: