Lucetta™ 2 Luminometer Manual

Summary of Contents for Lucetta 2

Page 1: ...Bioscience Solutions Lucetta 2 Luminometer Manual...

Page 2: ......

Page 3: ...ctions 5 2 2 Maintenance 5 2 3 Safety Instructions and Use Precautions 6 2 4 Storage Conditions 6 2 5 Batteries 6 2 6 Return Shipment 7 2 7 Instrument Disposal 7 3 0 Setup Instructions 7 3 1 Lucetta 2...

Page 4: ...ta Storage and PC Upload 11 5 1 Maximum Storage Capacity 11 5 2 Viewing Logged Measurements 11 5 3 Upload to a PC 11 6 0 Setting Instrument Parameters 12 6 1 Password 12 6 2 Customer Settings 12 6 3 D...

Page 5: ...abled by the Lucetta Connect software Broad range of applications Lucetta 2 Luminometer applications include MycoAlert PLUS and MycoAlert Mycoplasma Detection Assays ATP assays e g ViaLight Cell Proli...

Page 6: ...strument is not approved for use in fire or explosion potential areas nor for use with flammable or explosive media The sample holder must remain free of dust and contamination The holder should only...

Page 7: ...tance if required 2 7 Instrument Disposal Decontaminate the instrument before disposal This luminometer contains electronic parts To prevent environmental pollution please dispose of the instrument an...

Page 8: ...nd the age of the battery The battery may auto discharge when the instrument is not used for an extended period To optimize the battery lifespan the display light switches off after 15 sec if the inst...

Page 9: ...from 0 to 9 Press OK The instrument will first perform reading A for all samples and then continue with reading B for all samples For convenient handling of incubation times we recommend a maximum num...

Page 10: ...efine a new protocol 4 Each measurement is given an identifier number displayed under TEST By default the number begins with 1 counting upwards The number can be customized by using the arrow buttons...

Page 11: ...ns to select day month or year Depending on the selected country for the date format for more details see section 6 0 the format is MM DD YYYY USA or DD MM YYYY Europe Theflashingcursorindicatesthenum...

Page 12: ...word for a new instrument is 0000 Maximum number of digits for a password is four 1 Navigate down to SETTINGS and press OK 2 Select PASSWORD and press OK 3 Enter OLD PASSWORD To change the given numbe...

Page 13: ...Figure 13 Display for setting date and time 6 4 System Display brightness and contrast are adaptable to the lighting conditions 1 Navigate down to SETTINGS and press OK 2 Select SYSTEM and press OK 3...

Page 14: ...liance with operating conditions Background or measurement value too low 0 RLU s Detector or electronics damaged or incorrect instrument settings Service required please contact Lonza Unexpected or wr...

Page 15: ...asuring protocols Data storage Storage of the last 2000 results Hardware and Software Requirements for Data Upload Optional PC with industrial standard Microsoft Windows 7 32 64 bit or higher Microsof...

Page 16: ...22 75323 Bad Wildbach Germany and distributed by Lonza Cologne GmbH Nattermannallee 1 50829 Cologne Germany and Lonza Walkersville Inc 8830 Biggs Ford Road Walkersville MD 21793 Microsoft Windows and...