Avoiding Electrical Shock 

 

Follow the guidelines below for safe operation of the unit.   
The FlashGel™ Power Supply has been designed to utilize shielded wires thus 
minimizing potential shock hazard to the user. Lonza recommends against the use of 
unshielded wires. 
To reduce risk of electrical shock:   
1.  In the event of solution spilling on the instrument, it must be dried out for at least 2 

hours and restored to NORMAL CONDITION before each operation. 

2.  Never connect or disconnect wires loading from the power jacks when the red 

indicator light of power switch is on. 

3.  WAIT at least 5 seconds after stopping a run before handling output leads or any 

connected apparatus. 

4.  ALWAYS make sure that your hands, work area, and instruments are 

clean 

and 

dry

 

before making any connections or operating the power supply. 

5.  ONLY connect the power cord to a properly grounded AC outlet. 
 

Avoiding Damage to the Instrument   

1.  Do not attempt to operate the device if damage is suspected. 
2.  Protect this unit from physical damage, corrosive agents and extreme temperatures 

(direct sunlight, etc.). 

3.  For proper ventilation and safety concerns, keep at least 10 cm of space behind the 

instrument, and at least 5 cm of space on each side. 

4.  Do not operate the unit out of environmental conditions previously listed. 
5.  Prior to applying any cleaning or decontamination methods other than in accordance 

with manufacturer’s recommendation, users should check with the manufacturer’s 
instructions to see if the proposed method will damage the equipment. 

 

Equipment Operation   

Follow the guidelines below for safe operation of the unit: 
1.  NEVER access dangerous chemicals or other materials to reduce the likelihood of 

possible hazard of explosion and damage. 

2.  Do not operate the unit without lids or covers. 
3.  A temporary conductivity caused by condensation might occur even though this 

series is rated Pollution Degree 2 in accordance with IEC 664. 

 

 

Summary of Contents for FlashGel 57024

Page 1: ...FlashGel Power Supply Instruction Manual Lonza Rockland Inc www lonza com scientific support lonza com Scientific Support 800 521 0390 Customer Service 800 638 8174 Rockland ME 04841...

Page 2: ......

Page 3: ...1 Table of Contents Safety Information 2 Maintenance 3 Specifications 4 Introduction 7 Operation Instructions 9 Troubleshooting 12 Ordering Information 15 Warranty and Liability Information 17...

Page 4: ...powe tive termi re to adhe atory haza power so safety alw ible to una quipment 2 ion user to im or possible arning of po sult in deat posal instruc or utmost p please recy tement e serious in the oper...

Page 5: ...se contact a qualified service representative or scientific support lonza com Environmental Conditions Ensure the instrument is installed and operated strictly under the following conditions 75 kPa 10...

Page 6: ...roprocessor controller Timer 1 999 minutes with alarm continuous Safety Device No load detection shrouded plugs and sockets Physical Specifications Terminal Pairs 2 Pairs Display 3 digit LED Construct...

Page 7: ...e could void the user s authority to operate the equipment It is strongly recommended for the user to read the following points carefully before operating this equipment 1 Read and follow the manual i...

Page 8: ...ounded AC outlet Avoiding Damage to the Instrument 1 Do not attempt to operate the device if damage is suspected 2 Protect this unit from physical damage corrosive agents and extreme temperatures dire...

Page 9: ...d FlashGel Camera offers an all in one comprehensive package for fast easy nucleic acid separation Its compact size with two terminal pairs which provide sufficient and accurate output voltages can de...

Page 10: ...atures 2 electrode pairs allowing for up to 2 FlashGel Cassettes to be run simultaneously saving both time and valuable bench space With a universal voltage rating the FlashGel Power Supply is also de...

Page 11: ...cates your current parameter If you are setting a value for Voltage the Voltage LED will be lit b Operation Mode After you start the experiment the LED light next to the RUN STOP button will light up...

Page 12: ...the unit by pressing the ON OFF switch 4 Connect the DC output jacks from the FlashGel Dock or other electrophoresis unit insert the red lead into the red output jack and the black lead into the blac...

Page 13: ...ill light up 10 Press button again to stop the unit at any time if necessary 11 When the run is completed Press button to terminate a timed run and turn the AC power OFF by the switch on the rear 12 T...

Page 14: ...correct cords The fuse has blown Replace the fuse Repeated fuse broken Hardware failure Contact Lonza Scientific Support Operation stops Electrophoresis leads are not connected to the power supply or...

Page 15: ...if the problem persists Er3 Error message Thermal limitation reached Output voltage 10V 1 Check the connections 2 If Er3 error message persists the problem may be caused by internal fan failure Conta...

Page 16: ...ll flat blade screwdriver into the slot below the ON OFF switch Turn the screwdriver to gently pry open the fuse compartment NOTE The fuse compartment will not open with the power cord in place 3 Pull...

Page 17: ...y Pack Includes FlashGel Dock and FlashGel Power Supply 57065 FlashGel Camera Pack Includes FlashGel Dock and FlashGel Camera FlashGel Consumable Kits Cat No Description 57063 FlashGel DNA Kit Include...

Page 18: ...16 Notes...

Page 19: ...CIATED WITH THIS PRODUCT WHETHER BASED IN TORT CONTRACT STRICT LIABILITY OR ANY OTHER LEGAL THEORY IS EXPRESSLY LIMITED TO REPLACEMENT OF NONCONFORMING PRODUCT OR PAYMENT IN AN AMOUNT NOT TO EXCEED TH...

Page 20: ...Lonza Rockland Inc www lonza com scientific support lonza com Scientific Support 800 521 0390 Customer Service 800 638 8174 Document 00525706 0812 2 Rockland ME 04841...

Reviews: