13
Er1
Error message
Over current(400mA
limitation reached)
Check if the buffer concentration is
appropriate. Excessive buffer
concentration may cause over current
issue.* To clear the error message,
press the
START/STOP
button again
Er2
Error message
Over voltage(300V
limitation reached)
Press
START/STOP
button
to clear
the error message. Contact Lonza
Scientific Support if the problem
persists.
Er3
Error message
Thermal limitation
reached(Output
voltage <10V)
(1) Check the connections
(2) If
Er3
error message persists, the
problem may be caused by internal
fan failure. Contact Lonza Scientific
Support.
nLD
Message
No load is detected (1) Check the connections
(2) Check the buffer condition / buffer
level.
AL1
Alarm message
Max. watts(60W) of
power reached
Warning message for reference
*Not likely when used with FlashGel™ Dock
Encountering Problems
1. Check the troubleshooting section.
2. Call Scientific Support or e-mail