32
Min. & Max. Cell Size:
Average size of cells also may vary from cell type to cell type. With this
parameter, users can optimize the instrument to efficiently detect their own cells. The value can be
adjusted with 1 µm of increase per step.
Declustering Level:
Cultured mammalian cells may form clumps during culture or handling.
Declustering function of LUNA-
II™ provides efficient detection of clumped cells. LUNA-II™ provides
4 alternates: None, Low, Medium and High. This function is helpful to count sticky cells or rod-
shaped spores. High level declustering takes more time to analyze.