93
15. Select the server and perform an upgrade following the procedure in Section 7.2.3.
16. Save the System Upgrade Log for future reference and support
17. Restart FireSync Client.
18. Verify that settings are correct and all sensors are listed
19. Recalibrate the system
20. Exercise system functionality and report any issues
21. Now the system should be in the ready state for customer application to run
If the current installed Client is the same as the latest release, steps 9
and 10 may be skipped, but the remaining steps are still required to
unsure the proper Firmware is present on all other system components.
If a Station has been replaced, the Station settings can be restored from a previous
Server Backup after the system has been upgraded. See Section 8.1.