D9S001GE4-0101
52/
57
5.3.6.7
Calculating and saving the new sensor constant - sensor 3:
sensor 3:
EU version
: Press the button
0,5/0,1
𝒅𝒊𝒔𝒑𝒍𝒂𝒚𝒆𝒅 𝒗𝒂𝒍𝒖𝒆
𝒗𝒂𝒍𝒖𝒆 𝒐𝒇 𝒖𝒔𝒆𝒅 𝒘𝒆𝒊𝒈𝒉𝒕 (𝟖𝟎 𝒌𝒈)
× 𝒔𝒆𝒏𝒔𝒐𝒓 𝟑 𝒄𝒐𝒏𝒔𝒕𝒂𝒏𝒕 = 𝒏𝒆𝒘 𝒄𝒐𝒏𝒔𝒕𝒂𝒏𝒕
Write the new constant value
to the position of sensor 3
𝟖𝟕, 𝟓
𝟖𝟎
× 𝟏, 𝟗𝟓𝟑 = 𝟐, 𝟏𝟑𝟔
87,5
Summary of Contents for Wissner-Bosserhoff ELEGANZA 4
Page 5: ...D9S001GE4 0101 5 57 Identification of bed sides...
Page 18: ...D9S001GE4 0101 18 57 Weight module and tensometers S6015189ND...
Page 31: ...D9S001GE4 0101 31 57 i Drive Power connection scheme...
Page 34: ...D9S001GE4 0101 34 57 Castors Castors Antistatic castor Castor with lock Central Castor Control...
Page 35: ...D9S001GE4 0101 35 57 Bed Wiring Diagram...
Page 45: ...D9S001GE4 0101 45 57 The CAL position displays the calibration constant...
Page 49: ...D9S001GE4 0101 49 57 5 3 6 4 Positioning the weights for sensors 2 head left sensor...
Page 51: ...D9S001GE4 0101 51 57 5 3 6 6 Positioning the weights for sensor 3 left foot sensor...
Page 53: ...D9S001GE4 0101 53 57 5 3 6 8 Positioning the weights for sensor 4 right foot sensor...