5.3.5 Calibration constant setting
Before setting the calibration constant, the technician needs to set the local geographical zone. The technician can
set the final customer’s geographical zone after the scale calibration is finished.
When in Constant setting mode:
Go to calibration mode
– Simultaneously press
HOLD
,
CLEAR
,
ZERO
and
SPEAKER LEVEL
Press the
SPEAKER LEVEL
button and go to the
CAL
position
.
Press the
BED EXIT OUTER ZONE
button for 5 seconds
During this process the label ,,
ENTER
‘’ will display on the display
Now you are in calibration constant setting mode. The calibration setting is divided into four phases. The step
between is followed by sound signal “beep”.
5.3.5.1 Phase One
Display:
,,nC’’ (the number of calibrations) and the respective value are displayed alternately.
After balancing, the scale will move to the next phase.
5.3.5.2 Phase Two
,,L 0’’ (load 0) is displayed on the display. In this phase the bed is calibrated without weight.
Calibration:
Press the
SPEAKER LEVEL
button to start.
The calibration process takes approx. 7 seconds and the symbol ,,L
0’’ flashes the on the display.
The bed must be balanced throughout the calibration!
In this phase is set
absolute zero
value.
5.3.5.3 Phase Three
,,L160’’ (load 160) is displayed on the display. It is possible to change the weight level to 80 or 160. This will be
displayed on the display.
Calibration:
Change the weight level to 80 or 160 by pressing the
ZERO
button.
Place weights onto the bed according to the previous setting (80 kg or 160 kg).
Confirm the calibration by pressing the
SPEAKER LEVEL
button.
The calibration process will take approx. 7 seconds and the symbol ,,
L 160’’ or ,,L 80’’ flashes on the display.
The bed must be balanced throughout the calibration!
5.3.5.4 Phase Four
Display:
,,End’’ (directive correction ≤ 5%) and directive value are displayed alternately.
-or-
,,Err’’ (directive correction > 5%) and directive value are displayed alternately.
Result:
Press the
HOLD
button to save the value when the value is correct (,,End’’ is displayed). By this action the mode is
ended and the system will go back to the Constant setting mode.
-or-
Press the
SPEAKER LEVEL
button to go back to Phase One and repeat the process (,,Err’’ is displayed).
NOTE
The calibration mode can be cancelled at any time by pressing the CLEAR button for 5 seconds.
Scale service mode deactivation:
Press the
CLEAR
button
.
Summary of Contents for Wissner-Bosserhoff ELEGANZA 4
Page 5: ...D9S001GE4 0101 5 57 Identification of bed sides...
Page 18: ...D9S001GE4 0101 18 57 Weight module and tensometers S6015189ND...
Page 31: ...D9S001GE4 0101 31 57 i Drive Power connection scheme...
Page 34: ...D9S001GE4 0101 34 57 Castors Castors Antistatic castor Castor with lock Central Castor Control...
Page 35: ...D9S001GE4 0101 35 57 Bed Wiring Diagram...
Page 45: ...D9S001GE4 0101 45 57 The CAL position displays the calibration constant...
Page 49: ...D9S001GE4 0101 49 57 5 3 6 4 Positioning the weights for sensors 2 head left sensor...
Page 51: ...D9S001GE4 0101 51 57 5 3 6 6 Positioning the weights for sensor 3 left foot sensor...
Page 53: ...D9S001GE4 0101 53 57 5 3 6 8 Positioning the weights for sensor 4 right foot sensor...