LINET Wissner-Bosserhoff ELEGANZA 4 Service Manual Download Page 46

D9S001GE4-0101 

 
 

 

 

46/

57

 

 

 

5.3.5  Calibration constant setting 

Before setting the calibration constant, the technician needs to set the local geographical zone. The technician can 
set the final customer’s geographical zone after the scale calibration is finished. 
 

When in Constant setting mode: 

 

Go to calibration mode 

– Simultaneously press 

HOLD

CLEAR

ZERO

 and 

SPEAKER LEVEL 

 

Press the 

SPEAKER LEVEL

 button and go to the 

CAL 

position

.

 

 

Press the 

BED EXIT OUTER ZONE

 button for 5 seconds

 

 

During this process the label ,,

ENTER

‘’ will display on the display 

 

Now you are in calibration constant setting mode. The calibration setting is divided into four phases. The step 

between is followed by sound signal “beep”. 

5.3.5.1 Phase One 

Display: 

 

,,nC’’ (the number of calibrations) and the respective value are displayed alternately.

 

After balancing, the scale will move to the next phase. 

5.3.5.2 Phase Two 

,,L  0’’ (load 0) is displayed on the display. In this phase the bed is calibrated without weight. 

Calibration: 

 

Press the 

SPEAKER LEVEL

 button to start. 

The calibration process takes approx. 7 seconds and the symbol ,,L  

0’’ flashes the on the display.  

The bed must be balanced throughout the calibration! 

 

In this phase is set

 absolute zero 

value.

 

5.3.5.3 Phase Three 

,,L160’’ (load 160) is displayed on the display. It is possible to change the weight level to 80 or 160. This will be 
displayed on the display. 

Calibration: 

 

Change the weight level to 80 or 160 by pressing the 

ZERO 

button. 

 

Place weights onto the bed according to the previous setting (80 kg or 160 kg). 

 

Confirm the calibration by pressing the 

SPEAKER LEVEL

 button. 

The calibration process will take approx. 7 seconds and the symbol ,,

L  160’’ or ,,L  80’’ flashes on the display.  

The bed must be balanced throughout the calibration! 

5.3.5.4 Phase Four 

Display: 

 

,,End’’ (directive correction ≤ 5%) and directive value are displayed alternately.

 

-or- 

 

,,Err’’ (directive correction > 5%) and directive value are displayed alternately. 

Result: 

Press the 

HOLD

 

button to save the value when the value is correct (,,End’’ is displayed). By this action the mode is 

ended and the system will go back to the Constant setting mode. 

-or- 

Press the 

SPEAKER LEVEL

 

button to go back to Phase One and repeat the process (,,Err’’ is displayed). 

NOTE

   

The calibration mode can be cancelled at any time by pressing the CLEAR button for 5 seconds. 

 

 

Scale service mode deactivation: 

Press the 

CLEAR 

button

 

 

Summary of Contents for Wissner-Bosserhoff ELEGANZA 4

Page 1: ...D9S001GE4 0101 Version 03 Publication Date 2019 09 Service Manual ELEGANZA 4 Positionable Bed for Intensive Care version with scales and without scales...

Page 2: ...ers in foot side rails 13 Supervisor control panel ACP 15 Handset 15 Mobi Lift 15 3 7 1 Left and right side rail limit switches EMR ready Only 16 Weighing module 16 Weighing sensor tensometer 17 Integ...

Page 3: ...play 39 4 3 2 Stop codes 40 4 3 3 Operation codes 41 5 Bed Setup 42 Column and accelerometer calibration 42 5 1 1 Column and accelerometer calibration using the ACP 42 Transport lock 42 Scale Calibrat...

Page 4: ...echnical label description 1 Address of Distributor 2 Manufacturing Date Year Month Day 3 DI Device Identifier GTIN Global Trade Item Number 4 1D Bar code GS1 128 Serial Number 5 Symbols 6 Configurati...

Page 5: ...D9S001GE4 0101 5 57 Identification of bed sides...

Page 6: ...D9S001GE4 0101 6 57 2 Technical data and Symbols Follow User Manual for Eleganza 4 D9U001GE4 0101 to find Technical Data and used Symbols...

Page 7: ...and sends it to the control unit Mains power is supplied to the power box via a three core mains cable in the relevant national standard The power unit is standard at voltage 230V 50 60 Hz and altern...

Page 8: ...he leak is recorded simultaneously in the permanent memory of the control unit Positioning function locks and safety circuit to detect overload See Detecting overload 3 1 4 Technical parameters Parame...

Page 9: ...ccidental movement of the bed during a fault or from unintentional pressing of controllers 3 1 5 4 Simultaneous pressing of any two buttons The control unit is issued with the function for recognizing...

Page 10: ...ndition After an additional increase of approximately 15 C the switching power elements are disconnected this makes positioning impossible also signalled by a constant tone Information about the therm...

Page 11: ...e mode Information about charging and modes of the rechargeable battery is signalled by the rechargeable battery s LED as follows Off The rechargeable battery is charged one exception may be a certain...

Page 12: ...e panel is a production and technical label and the backrest accelerometer Due to absence of STOP button the bed must be equipped with ACP with STOP button 3 3 2 Backrest accelerometer in iBoard Basic...

Page 13: ...the control unit indicator lights for the mode and condition of the rechargeable battery a selection of special functions and settings for standard positions of the bed Foot siderails 3 4 1 Controller...

Page 14: ...D9S001GE4 0101 14 57 Number Description 1 iBoard Basic Cable Left Right 2 Foot Siderail Control Cable Siderail Cables 2 1 2 2 1 1...

Page 15: ...functions of both handsets are identical A cable with an 8 pin DIN connector is connected to a Plug Play adaptor which is linked to the control unit The controller has an RC unit to detect short circu...

Page 16: ...e side rail down Weighing module The weighing module allows the bed s mattress platform load to be calculated Four weighing sensors tensometers are connected to the weighing module The tensometer cons...

Page 17: ...lded cable The functions of the individual cores are described in the table Model number of tensometer 8300 0674A Model number of tensometer 8300 0674A Sensors T1 T2 PIN Meaning Colour Sensors T3 T4 P...

Page 18: ...D9S001GE4 0101 18 57 Weight module and tensometers S6015189ND...

Page 19: ...ls to the server Each device is marked with a label containing important information about the integration module The label contains the product number of the product power supply level type of hardwa...

Page 20: ...D9S001GE4 0101 20 57 EMR ready solution Mattress platform actuators Position Description 1 Backrest linear actuator 2 Thighrest linear actuator Mattress platform actuators 1 2...

Page 21: ...ble 1435 mm Compression strength 3 00 kN Tensile strength 0 kN Mechanical release Yes Input voltage 24 V Overvoltage protection No Cover IPX6 Max current load 3 A 3 12 2 Thighrest actuator The thigh m...

Page 22: ...olex connector data undercarriage module The head column is located near back rest part and foot column is located under the leg rest area The columns are fixed using two clamps to the frame undercarr...

Page 23: ...castor I Brake Voltage is measurable at the connector on pin number 1 and 3 see diagram connector No 8 The analogue signal of the UM allows the presence of mains power to the bed to be identified Thi...

Page 24: ...rriage frame with two screws 11 1 Nm There are three buttons on the controller lift upwards lift downwards and lift to examination position A cable runs from the foot controller to the undercarriage m...

Page 25: ...trol CCC replacing the activity of the operator The position of the CCC braked unbraked is detected by a single magnetic limit switch Its signal is transmitted via the undercarriage module to the cont...

Page 26: ...the bed from injuries due to crushing by the lowered Mattress Platform When obstacle occurs on the undercarriage and Mattress Platform is going down the motion is automatically stopped Safe Stop consi...

Page 27: ...d If the i Drive has been fully discharged the functions of the bed can still be used The bed has one i Drive Power controller 3 20 1 Technical specifications Specifications Value Diameter of i Drive...

Page 28: ...controller or the touch sensor is switched on The i Drive castor is controlled using the i Drive controller The controller contains four pressable buttons and one touch sensor If Forward and Reverse a...

Page 29: ...t is equipped with an input for a 36V rechargeable battery an AC charging source an input for the Braked Bed magnetic switch and an analogue motor temperature sensor input When the bed is plugged into...

Page 30: ...o that its lower voltage during gradual discharge is compensated for 3 20 8 i Drive rechargeable battery The driven castor rechargeable battery comprises three 12V rechargeable batteries connected in...

Page 31: ...D9S001GE4 0101 31 57 i Drive Power connection scheme...

Page 32: ...on in Control Unit Thermal Sensor Cable 13 Position in Control Unit Castor Control Position Cable with Sensor A i Brake Cable B i Drive Power Wheel Emergency Retraction Cable with switch C i Drive Pow...

Page 33: ...on 2 5th castor foot controled 3 Distance washer Fe Zn Cr d6 4 D18 T1 6 DIN 9021 4 Screw Fe Zn Cr M6 L20 DIN 7500C TORX 5 screw Fe Zn M6 L20 DIN 912 VAL Hex 6 Nut M 6 DIN 985 7 spring D9 T0 8 L80 DIN...

Page 34: ...D9S001GE4 0101 34 57 Castors Castors Antistatic castor Castor with lock Central Castor Control...

Page 35: ...D9S001GE4 0101 35 57 Bed Wiring Diagram...

Page 36: ...of electrical parts of bed Turns off after removed of the issue CU or press of the STOP button SAFE STOP Movement of the Mattress support platform stopped by function Safestop Remove an object from u...

Page 37: ...tery discharged completely Only CPR Blinking 100ms ON 100ms OFF Battery charging Blinking 200ms ON 1600ms OFF NOTE First value always means blinking of the LED and the second value means the LED is tu...

Page 38: ...faulty indicated on accumulator status indicator Flashes i Drive control box heat protection is activated 4 2 3 Fault signalisation i Drive Power The system is protected against failure states by stop...

Page 39: ...celerometer 0 80 Chair 8 AD STOP AD stop 100 120 1x STOP 65 85 2x STOP 9 OPER C Operation codes see 4 3 3 XX YY XX function code YY code of periphery 10 STOP C Bed stop code see 4 3 2 11 SYS ERR Syste...

Page 40: ...rature exceeded 7 column current overload 0 8 column current overload 1 10 actuator current overload backrest 11 actuator current overload thighrest 12 stopped due to resistance on the same function 1...

Page 41: ...eight adjustment 9 Lock pedals 10 Thighrest up 12 Thighrest down 13 Trendelenburg 15 Backrest up 16 CPR 17 Backrest down 18 Anti Trendelenburg 19 Mobilisation position 20 Height adjustment up 21 Cardi...

Page 42: ...nurse control panel Simultaneously press backrest up backrest down CPR and anti Trendelenburg buttons and hold for 3 seconds Check whether you are in calibration mode The lock functions and battery L...

Page 43: ...simultaneously for 5 seconds To exit diagnostic mode press CLEAR button 5 3 2 1 General instructions Display The identification number and respective value is displayed The first value displayed is a...

Page 44: ...esignated measuring instrument must be verified by an authorised person after any change of constants according to the metrological act 5 3 4 Calibration mode Press and hold the HOLD CLEAR ZERO and SP...

Page 45: ...D9S001GE4 0101 45 57 The CAL position displays the calibration constant...

Page 46: ...symbol L 0 flashes the on the display The bed must be balanced throughout the calibration In this phase is set absolute zero value 5 3 5 3 Phase Three L160 load 160 is displayed on the display It is p...

Page 47: ...5 it needs to calculate with actual displayed value 5 3 6 1 Location of the tensometers on the bed No 1 Head left No 2 Head right No 3 Foot left No 4 Foot right Zeroing the scale Zero display while t...

Page 48: ...D9S001GE4 0101 48 57 5 3 6 3 Calculating and saving the new sensor constant sensor 1 sensor 1 EU version Press the button 0 5 0 1 di Write the new constant value to the position of sensor 1 87 5...

Page 49: ...D9S001GE4 0101 49 57 5 3 6 4 Positioning the weights for sensors 2 head left sensor...

Page 50: ...D9S001GE4 0101 50 57 5 3 6 5 Calculating and saving the new sensor constant sensor 2 sensor 2 EU version Press the button 0 5 0 1 div Write the new constant value to the position of sensor 2 87 5...

Page 51: ...D9S001GE4 0101 51 57 5 3 6 6 Positioning the weights for sensor 3 left foot sensor...

Page 52: ...D9S001GE4 0101 52 57 5 3 6 7 Calculating and saving the new sensor constant sensor 3 sensor 3 EU version Press the button 0 5 0 1 Write the new constant value to the position of sensor 3 87 5...

Page 53: ...D9S001GE4 0101 53 57 5 3 6 8 Positioning the weights for sensor 4 right foot sensor...

Page 54: ...D9S001GE4 0101 54 57 5 3 6 9 Calculating and saving the new sensor constant sensor 4 sensor 4 EU version Press the button 0 5 0 1 Write the new constant value to the position of sensor 4 87 5...

Page 55: ...bration mode The calibration is divided into two phases 5 4 1 Phase One Display nC the number of calibrations and the respective value are displayed alternately After balancing the scale will move to...

Page 56: ...needed for servicing the bed Tools Socket set Screwdriver set Engineer s wrench set Hex key bit set Torx bit set Hex key set Knife Set of pliers Torque wrench Hammer nylon hammer Tape measure Crimping...

Page 57: ...n Date 2019 09 Copyright LINET s r o 2019 Translation LINET s r o 2019 All rights reserved Use this Service Manual in combination with User Manual for Eleganza 4 D9U001GE4 0101 All trademarks and bran...

Reviews: