D9S001GE4-0101
13/
57
3.3.3
Integrated Attendant control panel
It is connected to the control unit via a 14-core cable terminated by a 13-pin DIN connector. The ACP panel is
connected to the Plug&Play connector which is located under the head siderails. The ACP is fitted with STOP button.
The supervisor control panel contains a lock to block individual functions of the control unit, indicator lights for the
mode and condition of the rechargeable battery, a selection of special functions, and settings for standard positions of
the bed.
Foot siderails
3.4.1
Controllers in foot side rails
The control panels (left/right) allow the patient to
set the position
of the backrest and thighrest. Autocontour buttons
are also available.
Controller commands are detected as voltage increase via a cable connected to the control unit. The cables of the
side rail can be disconnected at the edge of the mattress platform using Mini Din connectors.
Summary of Contents for Wissner-Bosserhoff ELEGANZA 4
Page 5: ...D9S001GE4 0101 5 57 Identification of bed sides...
Page 18: ...D9S001GE4 0101 18 57 Weight module and tensometers S6015189ND...
Page 31: ...D9S001GE4 0101 31 57 i Drive Power connection scheme...
Page 34: ...D9S001GE4 0101 34 57 Castors Castors Antistatic castor Castor with lock Central Castor Control...
Page 35: ...D9S001GE4 0101 35 57 Bed Wiring Diagram...
Page 45: ...D9S001GE4 0101 45 57 The CAL position displays the calibration constant...
Page 49: ...D9S001GE4 0101 49 57 5 3 6 4 Positioning the weights for sensors 2 head left sensor...
Page 51: ...D9S001GE4 0101 51 57 5 3 6 6 Positioning the weights for sensor 3 left foot sensor...
Page 53: ...D9S001GE4 0101 53 57 5 3 6 8 Positioning the weights for sensor 4 right foot sensor...