Functioning
Check that all bed adjustments reach the maximum position.
If necessary, clean, lubricate or replace any worn spots and parts.
Electric Control
Plug connections:
Replace O-rings on connectors.
Check plug connections for dirt and defects.
Clean or replace if necessary.
Check that the plug connectors are properly seated.
Motors:
Check motor movement (adjust bed positions).
Check for incorrect and interrupted movements.
Have defective motors replaced if necessary.
Check cables for signs of wear and entanglement.
Install a new cable or have it replaced if necessary.
Battery:
Check that the battery is working properly (disconnect the bed from the mains).
Have the battery replaced if necessary.
Castors
Clean the castors completely.
Grease the castors if necessary.
Use Caro EP 2 by DEA or an equivalent grease.
Check that the castors work properly.
Forward Movement
Unrestricted Movement
Braked
Have the brakes adjusted if necessary.
Have any defective castors replaced.
Accessories
Check that all accessories (for example, lifting pole, siderails, infusion stand, etc.) are working properly.
Replace if necessary.