LINET Eleganza 3 User Manual And Technical Description Download Page 42

D9U001GZ5-0101_03 

 

 

42 

 

 

 
 
 
 

9.11.10 Writing Table

 

 

 

Fig. 20  

Writing Table

 

Writing table to be placed in the openings on the bed ends. The writing table is made of ABS. Maximum load 
3 kg. If necessary, the writing table can be fixated vertically. 

Summary of Contents for Eleganza 3

Page 1: ...User manual and technical description Eleganza 3 Positionable hospital bed for acute care D9U001GZ5 0101 Version 03 Publication date 2017 06...

Page 2: ...net com D9U001GZ5 0101 Version 03 Publication date 2017 06 Copyright LINET s r o 2017 Translation LINET s r o 2017 All rights reserved All trademarks and brands are the property of the appropriate own...

Page 3: ...and Regulations 16 5 Scope of Delivery and Bed Variants 17 6 Setup 18 6 1 Transport 18 6 2 Setup 18 7 Assembly 20 7 1 Bed Ends 21 7 2 Satellite Panel 21 7 3 Potential Equalisation 22 8 Operation 23 8...

Page 4: ...s against bodily injury Danger warns against potential mortal injury 1 1 2 Structure of Warning Notices Type and source of danger Measures to avoid the danger Signal words 1 2 Other Symbols 1 2 1 Inst...

Page 5: ...accidents due to electrical current Type B applied parts Safety isolating transformer general CE safety mark Jack for attachment of conductor for potential equalisa tion Safe working load Warning aga...

Page 6: ...D9U001GZ5 0101_03 6 1 4 Serial label...

Page 7: ...the manufacturer Do not apply excess weight or loads to the bed according to SWL During peak loads or unavoidable excess loads CPR place mattress platform in the lowest position Ensure that only one p...

Page 8: ...rd other serious injuries or damage the bed Ensure that the stipulated duty cycle on time is not exceeded Ensure that the moving parts of the bed are not blocked To prevent failures use the manufactur...

Page 9: ...lations Disconnect the bed from the mains during exceptional cases i e a storm 2 3 Risk Assessment Risk of injury or death due to use of incorrect equipment Always conduct the risk assessments require...

Page 10: ...care bed is suitable for Patients with weight 40 kg with height 146 cm with BMI 17 with a weight of up to 185 kg Personnel qualified medical staff any person familiar with the manual patient conditio...

Page 11: ...0 Maximum Thigh Rest Angle 30 Trendelenburg Anti Trendelenburg Position 16 16 Weight Depending on Equipment 138 kg Safe Workload including Mattress and Accessories 250 kg Max Lifting Pole Load 75 kg E...

Page 12: ...only The high frequency radiations are very low and unlikely to cause any interference to nearby electronic devices High frequency radiation CISPR 11 Class B Eleganza 3 is suitable for all institutio...

Page 13: ...spital environment Short time voltage drop short duration interruptions and slow voltage changes on the feeder input line IEC 61000 4 11 5 UT 95 short duration drop of UT within 0 5 cycles 40 UT 60 sh...

Page 14: ...icinity of the instrument marked with the following symbol Radiated high frequency phenomena IEC 61000 4 3 3 V m 80 MHz to 2 5 GHz 3 V m a It is not possible to accurately indicate field intensities f...

Page 15: ...ng to frequency of transmitter m Rated maximum output of transmitter W 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 40 0 40 0 1 0 37 1 26 1 26 1 1 17 4 00 4 00 10 3 69 12 65 12 6...

Page 16: ...e following standards EN 60601 1 2006 A1 2013 EN 60601 1 2 2007 EN 60601 2 52 2010 EN 60601 1 6 2010 EN ISO 10993 93 42 EEC 2011 65 EU The manufacturer adheres to a certified quality management system...

Page 17: ...th fifth castor Tente Integral 150 mm Tente Integral 150 mm with fifth castors Tente Integral 150 mm double castor Tente Integral 150 mm double castors with fifth castor Control Elements Supervisor pa...

Page 18: ...is activated Ensure that the brakes are released while moving the bed 6 2 Setup Set up the bed as follows Unpack the bed Check the delivery see Scope of Delivery and Bed Variants Install equipment an...

Page 19: ...mains control box 1 by pulling strap 2 Check if isolating foil is complete and undamaged as shown in Fig 2 If isolating foil is damaged contact the manufacturer s service department immediately 2 NOTE...

Page 20: ...f the Hospital Bed 1 Removable Foot Board 2 Supervisor Panel Nurse Control Panel 3 Eleganza Protector 4 Split Siderail Middle Section 5 Four part Mattress Platform with Ergoframe System 6 Mobi Lift ha...

Page 21: ...ings 1 Pull bed ends from sleeve fittings Lock sleeve fittings 2 Install the bed ends as follows Unlock sleeve fittings 1 Slide bed ends into sleeve fittings Lock sleeve fittings 2 7 2 Satellite Panel...

Page 22: ...D9U001GZ5 0101_03 22 7 3 Potential Equalisation The bed contains a protective conductor between all metal parts Fig 3 Ground Pin Connect the ground wire to the ground pin under the head board...

Page 23: ...and cold places from 10 C to 0 C 3 To prevent accumulators from being in the sunshine NOTE Batteries will remain in fully functional condition only for a certain period of time which is dictated by th...

Page 24: ...oken or battery fuses are faulty contact service department of the manufacturer in case of such signalisation Replacing the battery Damage to the bed due to incorrect battery replacement Have the batt...

Page 25: ...tatus will be cancelled automatically when the bed switches to sleep mode To cancel this status Press STOP button 8 3 Removing the Bed from Service Remove the bed from service as follows Disconnect th...

Page 26: ...Secure or remove any items on the bed The bed is operated by different control elements Control elements depending on the model Supervisor Panel Nurse Control Panel Handset Foot Control Satellite Pane...

Page 27: ...tor 6 Backrest Adjustment Button 7 Backrest Lock Button 8 Backrest Lock Indicator 9 Height Adjustment Buttons 10 Height Tilt Lock Button 11 Height Tilt Lock Indicator 12 Auto Contour Button 13 Foot Co...

Page 28: ...h function keys 3 6 9 and 12 NOTE If you press two function buttons at the same time the controller will recognise this as an error The controller immediately interrupts all bed movements Set the posi...

Page 29: ...30 seconds from any mattress platform position You cannot disable this function Cardiac Chair Position The cardiac chair position is for patients with cardiac arrhythmia and breathing difficulties Se...

Page 30: ...into the CPR position within 30 seconds from any mattress platform position You cannot disable this function NOTE For rapid mechanical positioning see CPR Backrest Release 9 2 Handset A handset is inc...

Page 31: ...g replacing a defective handset using the handset for another bed 9 3 Satellite Panel The satellite panel is optional The satellite panel is attached to the mattress platform frame behind the backrest...

Page 32: ...nted Activation 4 Raise Mattress Platform Foot Control Set the position as follows Press and hold the foot switch until the desired position is reached 9 5 Integrated Control Element Siderail The inte...

Page 33: ...llows Pull and hold release handle 1 Press the backrest down Mechanical Adjustment of Calf Rest Incorrect lowering can injure the patient Hold the calf rest by both handles at all times when lowering...

Page 34: ...he safety catches frame by 90 and push them in To shorten the bed Pull out the safety catches 1 at the foot end on both sides of the frame Turn the safety catches 1 by 90 The safety lock is released P...

Page 35: ...ails To operate the siderails Fig 11 Fold up the Split Siderail Risk of injury due to patient falling out of bed Ensure that the folded up siderails are securely anchored To check the stability press...

Page 36: ...nism 2 Dismount Eleganza Protector as follows Pull up the locking mechanism 2 of Eleganza Protector Pull the tube of Eleganza Protector out of the sleeve fitting on the corner of the foot board 1 Caut...

Page 37: ...castors The brakes are at the foot end of the undercarriage Castor control a Braked All of the castors are braked b Unrestricted Movement All of the castors are unlocked c Forward Movement The front...

Page 38: ...serted in the sleeve fittings Ensure that no bed linen is caught between the sleeve fitting and the support handle To adjust the support handle Pull the support handle out from under the mattress plat...

Page 39: ...y pin locks into place A plastic grab handle with an adjustable strap shall be attached to the lifting pole NOTE The date of manufacture is marked on the grab handle LINET recommends that you replace...

Page 40: ...r hook is 2 kg 9 11 3 Accessory Rails The bed is optionally equipped with two accessory rails with plastic hooks Load Capacity Maximum load of 5 kg without leverage Maximum load of hook pair 10 kg Acc...

Page 41: ...ff as well as the patient to orientate themselves at night Requirements for Operation The main room lights must be dimmed or off Ensure that the bed is connected to the mains NOTE The night light is t...

Page 42: ...03 42 9 11 10 Writing Table Fig 20 Writing Table Writing table to be placed in the openings on the bed ends The writing table is made of ABS Maximum load 3 kg If necessary the writing table can be fix...

Page 43: ...orrosive or caustic detergents Never use detergents that deposit calcium carbonate Never use detergents with solvents that might affect the structure and consistency of the plastics benzene toluene ac...

Page 44: ...ng Patients Clean the following bed parts All of the control elements for adjusting the bed All handles Back and calf rest handles CPR release handle Head and foot boards Siderails in highest position...

Page 45: ...rds Siderails in highest position Freely accessible mattress surface Mobi Lift Accessory rails All plastic mattress platform covers Plastic undercarriage covers Telescopic columns Mattress on all side...

Page 46: ...ive battery Check the mains connection Notify the service department Plug inserted incorrectly Insert the mains plug correctly Faulty power source Notify the service department Faulty control element...

Page 47: ...manufacturer or service organisation immediately 12 1Maintenance Work Ensure that the following maintenance work is performed every 12 months by the manufacturer or by a qualified service organisatio...

Page 48: ...tors replaced if necessary Check cables for signs of wear and entanglement Install a new cable or have it replaced if necessary Battery Check that the battery is working properly disconnect the bed fr...

Page 49: ...service organisations in which the manufacturer declares that the service organisation is qualified to perform maintenance on LINET products LINET makes every reasonable effort within research design...

Page 50: ...kaging material according to the symbols and by delivering it to an authorised person The product consists of recyclable steel plastic and electronic components 13 2 Disposal Within Europe To dispose...

Page 51: ...Do not continue to use the bed This product is covered by a 24 month warranty from the date of purchase The warranty covers all material and manufacturing related failures and errors Failures and err...

Page 52: ...D9U001GZ5 0101_03 52 15 EC Declaration of Conformity...