D9U001GZ5-0101_03
18
6
Setup
6.1
Transport
To avoid damage during transport, please observe the following guidelines:
Ensure that no cables are run over when moving a bed.
Ensure that the mains cable is attached with a hook (at the head end of the bed).
Ensure that the castors are unlocked before moving the bed during the loading/unloading process
(see Undercarriage).
Move the bed on suitable floor surfaces only.
Suitable surfaces:
Tile
Hard linoleum
Poured flooring
Unsuitable surfaces:
Too soft, unsealed or defective flooring
Soft wooden flooring
Soft and porous stone floors
Carpeted floors with underlay
Soft linoleum
For longer distances, ensure that the castor steering function (main control) or fifth castor (separate
control) is activated.
Ensure that the brakes are released while moving the bed.
6.2
Setup
Set up the bed as follows:
Unpack the bed.
Check the delivery (see Scope of Delivery and Bed Variants).
Install equipment and accessories (see Assembly).
In case of delivery with dismantled ends, mount the head and foot ends (see Bed Ends).
Set up the bed on a suitable floor surface only.
Suitable surfaces:
Tile
Hard linoleum
Poured flooring
Unsuitable surfaces:
Too soft, unsealed or defective flooring
Soft wooden flooring
Soft and porous stone floors
Carpeted floors with underlay
Soft linoleum
Ensure that the mains cable does not collide or get stretched when adjusting the bed.
Check that the plug is inserted correctly.
Do not leave any extension cords or power strips loose on the floor.
Ensure that all of the required mechanical and electrical prevention mechanisms are available onsite
(protective wall bumpers, corner bumpers, circuit breakers, etc.).