D9U001GZ5-0101_03
47
12
Maintenance
Warning
Risk of injury when working on the bed!
Ensure that the bed is disconnected from the mains connection prior to assembly,
disassembly and maintenance.
Ensure that the castors are locked prior to assembly, disassembly and
maintenance.
A defective bed can cause injuries!
Have a defective bed repaired immediately.
If the defect cannot be repaired, do not use the bed.
Warning
Caution
Incorrect maintenance can damage the bed!
Ensure that maintenance is performed by customer service or trained hospital
technicians only.
Do not use the bed if any malfunction or defect occurs. In this case contact
manufacturer or service organisation immediately.
12.1 Maintenance Work
Ensure that the following maintenance work is performed every 12 months by the manufacturer or by a
qualified service organisation trained and certified by the manufacturer
Spare Parts
The product label is located on the inside of the longitudinal rail of the mattress platform frame. The product label
contains information for claims and ordering replacement parts.
Information about spare parts is available from:
Customer service
Sales
Our technical support department
Completeness
Perform a visual check (with delivery note if necessary).
Have any missing parts replaced.
Wear
Check all bolts and tighten if necessary.
Check all locking mechanisms.
Check the bed for wear, scratches or rub marks.
Eliminate the cause if necessary.
Have any defective parts replaced.