MODBUS Register Map v1.48
248083408-1 Rev 1
A-1
A
MODBUS Register
Map v1.48
COMM
Settings
Lighthouse particle counters with MODBUS have the following
communications settings:
The MODBUS slave address is set on the particle counter. Valid
addresses are 1-63. Address 0 is the broadcast address.
Supported
MODBUS
Commands
See www.modbus.org for documentation on how to use these
commands.
Table A-1 MODBUS Communications Settings
Baud Rate
19200
Data Bits
8
Stop Bits
1
Parity
None
Hardware Protocol
RS-232C or RS-485 Standard
Software Protocol
MODBUS ASCII (supports upper/lower case)
Table A-2 MODBUS Registers
Hex Command
Description
03
Read Holding Registers
04
Read Input Registers
06
Write Single Holding Register
Summary of Contents for Remote 2014P
Page 1: ......
Page 2: ......
Page 6: ......
Page 10: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual t iv 248083408 1 Rev 1...
Page 12: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual ii 248083408 1 Rev 1...
Page 14: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual 1 2 248083408 1 Rev 1...
Page 20: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual 2 6 248083408 1 Rev 1...
Page 40: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual 3 20 248083408 1 Rev 1...
Page 72: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual B 2 248083408 1 Rev 1...
Page 74: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual C 2 248083408 1 Rev 1...
Page 79: ......
Page 80: ......