Maintenance Procedures
248083408-1 Rev 1
5-1
5
Maintenance
Procedures
This Chapter provides the user with the procedures to perform in
maintaining the REMOTE ’P’ instruments.
Safety
Before performing any of the maintenance tasks described in this
chapter, review the safety warnings supplied throughout this manual.
Calibration
To maintain optimum performance of this instrument, it should be
recalibrated annually by a Lighthouse Authorized Service Provider.
Cleaning
This procedure may be superceded by customer requirements;
however, not under any circumstances apply Acetone to the REMOTE
’P’ instrument.
1.
To keep the interior of the laser sensor clean, remove the
isokinetic probe from the air inlet and install the protective plastic
cap supplied with the instrument.
2.
Moisten a lint-free cloth with isopropyl alcohol. The cloth should
be moist, not wet.
3.
Wipe down the exterior surfaces of the instrument.
Purge Test
1.
Connect the Purge filter to the sample inlet. The purge filter
should be a 0.1 micron, 0.1 CFM filter.
2.
Apply power to the instrument.
3.
Configure the unit via the user interface to sample for 30 minutes.
4.
Allow the instrument to sample through a 30 minute period. This
time allows the unit to warm up and purge any residual particles
that might be inside it.
Summary of Contents for Remote 2014P
Page 1: ......
Page 2: ......
Page 6: ......
Page 10: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual t iv 248083408 1 Rev 1...
Page 12: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual ii 248083408 1 Rev 1...
Page 14: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual 1 2 248083408 1 Rev 1...
Page 20: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual 2 6 248083408 1 Rev 1...
Page 40: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual 3 20 248083408 1 Rev 1...
Page 72: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual B 2 248083408 1 Rev 1...
Page 74: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual C 2 248083408 1 Rev 1...
Page 79: ......
Page 80: ......