32
In the following pages, the different parameters are explained and the standard settings are given.
Parameters that are not used yet are not accessible or shown with underscores.
PARAMETERS:
Parameter
Function
Settings
Default
US
01
Start-up unit (and print units)
1=kg / 2=lb.
2
02
Smallest graduation step for
multi-range
0.1/0.2/0.5…………10/20/50
0.5
03
Largest graduation step for
multi-range
0.1/0.2/0.5 ………...10/20/50
2
04
Number of graduations for
every range
0000-9900 divisions
1000
05
Weighing capacity system (full
scale)
0000-99999 units
5000
06
Motion tolerance for stable
0-32
1
off 0.5 grad./sec
1 grad./sec 2 grad./sec
4 grad./sec 8 grad./sec
16 grad./sec 32 grad./sec
07
Filter size
0-12
8
0=off
1=light filtering, 12=heavy filtering
08
Auto zero range
0=off 0.5 division
0.5
1=division 3 divisions
09
Zero range positive (+)
0-100% (approved 2%) of span
10
10
Zero range negative (-)
0-100% (approved 2%) of span
10
11
Test Function
BASIC ADC Counts
BASIC
10x Resolution
12
Not used
13
Not used
14
Start-up number to add in
sampling mode
1-2-5-10-20-50-95
10
15
Units switch mode active
Yes / No
Yes