ACT I Sensor Technical Specifications
Parameter
Min
Max
Typical
Units
Input Impedance
19.5
20.5
20
MΩ
Input dynamic range
+/- 5
mV
Average current
consumption
4
12
6.5
mA
Peak current
Consumption
N/A
80
75
mA
CMRR
75
dB
ADC sample rate
250
250
250
samples/sec
DC offset correction
150
|mV|
LPF cutoff frequency
61.8
90.45
76
Hz
HPF cutoff frequency
.043
.065
.005
Hz
Impedance
measurement range
0
793
KΩ
Arrhythmia algorithm
detection voltage range
0.4
5
|mV|
System noise
32
μV
Pacemaker pulse width
marking
0.2
2
msec
Pacemaker pulse
amplitude marking
2
250
mV
ECG data buffer
32
MBit
ECG buffering time
2:19
Hrs
Manual ECG event
triggering
Yes
PM detection
Yes