39
RESULTS
◆
Factors such as a low or high hematocrit
(the percentage volume of red blood
cells in your blood). For example, if your
hematocrit is too low (below
25
%) or too
high (above
60
%).
◆
Severe dehydration (loss of body fluid) may
also cause meter results to be different
from a laboratory result. Repeated vomiting
or diarrhea, both of which can occur when
your glucose is very high, can cause this.
Your blood may become thick, with not
enough liquid to properly wet the test strip.
This may cause your meter to give false low
blood glucose results.
◆
If you test at a very low or high tempera-
ture or a low or high relative humidity or
some other extreme environmental con-
dition. If you test under these conditions,
repeat the meter test in a climate-
controlled place as soon as possible. The
operating temperature is
10
–
35
°C (
50
–
95
°F). The operating humid-
ity is
10
–
90
%. You should always test
within these ranges.
◆
You may also get inaccurate results if
you test when both the temperature and
relative humidity are near the limits
defined above. If you test under these
conditions, repeat the meter test in a
climate-controlled place as soon as
possible.
◆
Conditions affecting circulation to the
hands may also lead to inaccurate
results. These include shock, some
drugs called vasoactive agents, and
Raynaud’s disease.
4
,
7
,
8
See the
Important Technical and Healthcare
Professional Information
section in the
OneTouch
®
SureStep
®
Test Strip package
insert for typical accuracy and precision data
for the OneTouch
®
SureStep
®
system. Make
DCO_05090602A_OTSS_OB_CA_ENG 12/3/03 1:27 PM Page 39
Summary of Contents for Blood Glucose Meter
Page 9: ...DCO_05090602A_OTSS_OB_CA_ENG 12 3 03 1 27 PM Page viii...
Page 10: ...DCO_05090602A_OTSS_OB_CA_ENG 12 3 03 1 27 PM Page 1...
Page 85: ...DCO_05090602A_OTSS_OB_CA_ENG 12 3 03 1 27 PM Page 76...
Page 94: ...DCO_05090602A_OTSS_OB_CA_FR 12 3 03 1 31 PM Page viii...
Page 95: ...DCO_05090602A_OTSS_OB_CA_FR 12 3 03 1 31 PM Page 1...
Page 170: ...DCO_05090602A_OTSS_OB_CA_FR 12 3 03 1 31 PM Page 76...