
21
CHECKING
◆
you are not following all test steps or
performing them incorrectly.
◆
the code number on the meter does not
match the code number on the test
strip vial.
◆
the meter or test strip holder is dirty.
◆
the control solution is expired, or beyond
its discard date, contaminated, or not
well mixed.
◆
there is a problem with the test strip
(expired, damaged, etc.).
◆
the meter has malfunctioned.
◆
the control solution test is performed
outside of the temperature range
15
–
35°C (59–95°F)
.
◆
you have applied too much control
solution.
If your control solution result falls outside
the printed range, repeat the control solu-
tion test with a new test strip. If your second
control solution result is also outside the
printed range, call the OneTouch
®
Customer
Care Line at
1 800 663-5521.
WARNING:
Do NOT use your system to test
your blood when your control solution
results fall outside the printed range. You
may get inaccurate blood glucose results.
Call the OneTouch
®
Customer Care Line
at
1 800 663
–
5521
if this occurs.
DCO_05090602A_OTSS_OB_CA_ENG 12/3/03 1:27 PM Page 21
Summary of Contents for Blood Glucose Meter
Page 9: ...DCO_05090602A_OTSS_OB_CA_ENG 12 3 03 1 27 PM Page viii...
Page 10: ...DCO_05090602A_OTSS_OB_CA_ENG 12 3 03 1 27 PM Page 1...
Page 85: ...DCO_05090602A_OTSS_OB_CA_ENG 12 3 03 1 27 PM Page 76...
Page 94: ...DCO_05090602A_OTSS_OB_CA_FR 12 3 03 1 31 PM Page viii...
Page 95: ...DCO_05090602A_OTSS_OB_CA_FR 12 3 03 1 31 PM Page 1...
Page 170: ...DCO_05090602A_OTSS_OB_CA_FR 12 3 03 1 31 PM Page 76...