StoreX
Series
Remote Operation V0712 .doc
28
RKi, 26.11.2020
DISCONTINUED
2.1.6
Random Access Cycles
Program for positioning the carrousel randomly including gate movements.
After entering the access intervals the program continuously simulates accesses at random
positions until the “q-Key“ is pressed. As a result a table containing statistical data is
presented.
If an error of positioning time larger then 0.4 seconds compared with the calibrated
positioning time is observed a positioning error is assumed and monitored.
2.1.7
Random Fast Access
Program for positioning the carrousel randomly including combined positioning-gate
movements commands.
After entering the access intervals the program continuously simulates accesses at random
positions until the “q-Key“ is pressed. As a result a table containing statistical data is
presented.
If an error of positioning time larger then 0.4 seconds compared with the calibrated
positioning time is observed a positioning error is assumed and monitored.