background image

 

StoreX

 

Series

  

 

Remote Operation V0712 .doc 

20

  RKi, 26.11.2020 

DISCONTINUED 

with manual rotation of carrousel 

Stacker Slot Error 

Stacker slot cannot be reached 

00011 

cr lf

 

Remote Access Level Error 

Undefined stacker level has been 
requested 

00012 

cr lf

 

Plate Transfer Detection 
Error 

Export operation while plate is on transfer 
station 

00013 

cr lf

 

Lift Initialization Error 

Lift could not be initialized 

00014 

cr lf

 

Plate on Shovel Detection 

Trying to load a plate, when a plate is 
already on the shovel 

00015 

cr lf

 

No Plate on Shovel Detection  Trying to remove or place plate with no 

plate on the shovel 

00016 

cr lf

 

No recovery 

Recovery was not possible 

00017 

cr lf

 

 
 
 
 

Error 

Description 

Code 

Import Plate Stacker  
Positioning Error 
 

Carrousel could not reach desired radial 
position during Import Plate procedure or 
Lift could not reach transfer level during 
Import Plate procedure. 

00100 

cr lf

 

Import Plate Handler Transfer 
Turn out Error 
 

Handler could not reach outer turn position 
at transfer level during Import Plate 
procedure. 

00101 

cr lf

 

Import Plate Shovel Transfer 
Outer Error 
 

Shovel could not reach outer position at 
transfer level during Import Plate 
procedure. 

00102 

cr lf

 

Import Plate Lift Transfer 
Error 

Lift did not reach upper pick position at 
transfer level during Import Plate 
procedure. 

00103 

cr lf

 

Import Plate Shovel Transfer 
Inner Error 

Shovel could not reach inner position at 
transfer level during Import Plate 
procedure. 

00104 

cr lf

 

Import Plate Handler Transfer 
Turn in Error 

Handler could not reach inner turn position 
at transfer level during Import Plate 
procedure. 

00105 

cr lf

 

Import Plate Lift Stacker 
Travel Error 

Lift could not reach desired stacker level 
during Import Plate procedure. 

00106 

cr lf

 

Import Plate Shovel Stacker 
Front Error 
 

Shovel could not reach front position on 
stacker access during Plate Import 
procedure. 

00107 

cr lf

 

Import Plate Lift Stacker 
Place Error 

Lift could not reach stacker place level 
during Import Plate procedure. 

00108 

cr lf

 

Import Plate Shovel Stacker 
Inner Error 

Shovel could not reach inner position at 
stacker plate placement during Import Plate 
procedure. 

00109 

cr lf

 

Import Plate Lift Travel Back 
Error 

Lift could not reach zero level during Import 
Plate procedure. 

00110 

cr lf

 

Import Plate Lift Init Error 
 

Lift could not be initialized after Import 
Plate procedure. 

00111 

cr lf

 

 

Summary of Contents for StoreX BT

Page 1: ...e Operation for StoreX Precision Incubator Series DISCONTINUATION This Interface is no longer supported Do not use for new development STX40 STX100 STX200 STX500 STX1000 IC HC HR DC DR DF SA BT IT STP...

Page 2: ...ot follow the instructions given in this manual LiCONiC does not take any responsibility for injuries or damages caused by the LiCONiC product Contact Address LiCONiC AG Industriestrasse 8 FL 9493 Mau...

Page 3: ...RKi 26 11 2020 DISCONTINUED Table of Contents 1 Remote Operation 5 1 1 RS 232 Serial Port Configuration 5 1 2 Command Transmission Procedure 5 1 3 Commands 11 1 4 Handling Status Error Messages 19 1 5...

Page 4: ...ies Remote Operation V0712 doc 4 RKi 26 11 2020 DISCONTINUED History Page Document Change History Date Version Changes Author 26 11 2020 V7 12 Discontinuation This interface is not Supported anymore R...

Page 5: ...al port is reset by sending a Brake Signal The specification of the Brake Signal is given below In most cases the Brake Signal can be omitted The example program shows how a Brake Signal can be implem...

Page 6: ...mand PC Controller Resonse PC Controller 1 2 5 Command Syntax For communication only a few commands are required A command is an ASCII string which is sent to the controller Response is an ASCII strin...

Page 7: ...ackStr pN INTEGER s STRING STRING CONST tries 2 VAR i n m err INTEGER w WORD s0 s1 STRING c kp CHAR BEGIN IF NOT kbdEsc THEN BEGIN EmptyAux pN i 1 s0 s m Pos s0 IF m 0 THEN BEGIN Delete s0 m 1 s1 Copy...

Page 8: ...NG BEGIN Str rel sR STX_Set STX_ReadBackStr pN ST sR END FUNCTION KV_Reset pN rel INTEGER STRING VAR sR STRING BEGIN Str rel sR STX_Reset STX_ReadBackStr pN RS sR END FUNCTION STX_Read pN rel INTEGER...

Page 9: ...lf Write Protected Error Unauthorized Access E4 cr lf Base Unit Error Unauthorized Access E5 cr lf 1 2 8 List of Operations Operations are physically executable commands Operations are triggered by on...

Page 10: ...ready to accept commands or 1 handling ready command may be sent In its idle status the handling sets the Ready Bit to 1 Command Response Read Ready Bit RD sp 1915 cr x cr lf When polling the Ready Bi...

Page 11: ...5 sp n cr OK cr lf Set Handler Stacker Pitch WR sp DM23 sp p cr OK cr lf Open Gate2 ST sp 1901 cr OK cr lf Close Gate3 ST sp 1902 cr OK cr lf Terminate Access ST sp 1903 cr OK cr lf Import plate to m...

Page 12: ...ified slot Note that stacker levels are numbered starting at the bottom level upwards Bottom level is 1 The Activate Handler Command initializes the handling Use the Activate Handler Command after col...

Page 13: ...OK cr lf Position rotation at slot 2 WR sp DM5 sp 10 cr OK cr lf Select level 10 in stacker ST sp 1904 cr OK cr lf Import Command to start import process To export a plate form level 22 in the stacke...

Page 14: ...o Way Communication Activation Flag default 0 RS sp 1200 cr OK cr lf Read Flag Send Handling Progress Status SPX default 0 RD sp 1201 cr x cr lf Set Flag Send Handling Progress Status SPX default 0 ST...

Page 15: ...1611 cr OK cr lf Read Wait Plate selection RD sp 1612 cr x cr lf Wait Plate on stop at x fer station until plate present resp x fer station is free ST sp 1612 cr OK cr lf Wait Plate off do not check...

Page 16: ...Stacker position default 940 RD DM81 cr ddddd cr lf Read Handler Transfer Station position default 3500 RD DM82 cr ddddd cr lf Set Handler z Offset WR DM20 sp d cr OK cr lf Set Handler dz Pick Palce M...

Page 17: ...M for plate transfer Daughter Plate ST 1105 cr OK cr lf Select upper level on TPM for plate transfer Mother Plate RS 1105 cr OK cr lf Read lift transfer height for plate transfer at TPM RD DM41 cr ddd...

Page 18: ...ated in DM251 DM499 DM251 represents cassette location number 1 DM252 cassette location number 2 A total of 250 cassettes can be configured Each entry of the Cassette Configuration Table contains the...

Page 19: ...s stacked and executed after termination of the first command 1 4 Handling Status Error Messages Error Commands are used when the handling detects an internal error e g when loading a plate to an occu...

Page 20: ...er Error Shovel could not reach outer position at transfer level during Import Plate procedure 00102 cr lf Import Plate Lift Transfer Error Lift did not reach upper pick position at transfer level dur...

Page 21: ...r place position at transfer level during Export Plate procedure 00207 cr lf Export Plate Shovel Transfer Inner Error Shovel could not reach inner position at transfer level during Export Plate proced...

Page 22: ...EDURE STX_WaitReadyTrace portNbr INTEGER VAR c CHAR x y err INTEGER n WORD BEGIN WRITE x WhereX y WhereY DelayMs 300 WHILE STX_Read prtNbr 1915 0 AND NOT KeyPressed DO BEGIN DelayMs 100 GotoXY x y Val...

Page 23: ...StoreX Series Remote Operation V0712 doc 23 RKi 26 11 2020 DISCONTINUED...

Page 24: ...or most StoreX commands is given below Use this procedure after initializing the instrument only PROCEDURE STX_DoPlate pN slot level INTEGER command STRING BEGIN IF NOT KeyPressed THEN BEGIN STX_WaitR...

Page 25: ...S Nbr Direct Commands 1 Monitor Flags 2 Macros 3 Teach Positioning Times 4 Random Positioning 5 Random Access Cycles 6 Random Fast Access 7 Quit 0 ____________________________________________________...

Page 26: ...message 0 means inactive the status message 1 means active Code Status Flag Comment P0 0000 0 initiator of carrousel Pn 0001 Position initiator of carrousel GtClse 0002 Gate closed switch GtOpen 0003...

Page 27: ...d R Reset Handling Controller Q Exit macro menu Accs Carrousel access remote or manual Ready Ready Bit 2 1 4 Teach Positioning Times Program to measure positioning times of the carrousel The program s...

Page 28: ...r then 0 4 seconds compared with the calibrated positioning time is observed a positioning error is assumed and monitored 2 1 7 Random Fast Access Program for positioning the carrousel randomly includ...

Reviews: