8) Connect the power supply to the BioStack Plate Stacker and tighten the threaded retainer ring
until it is finger tight. Plug the power cord into a wall outlet.
9) Turn on the BioStack Plate Stacker. Initialization takes about 20 seconds.
10) If necessary, connect the power and network cables to the Odyssey Sa Imager as described earlier
in this manual.
Important:
The BioStack™ must always be turned on before the Odyssey Sa Imager.
11) Turn on the Odyssey Sa Imager. The Odyssey Sa Imager should be turned on for all remaining
steps.
Leveling the Odyssey Sa Imager
12) If necessary, move the BioStack Plate Stacker so there is at least 18 cm (7 in.) clearance between
the Odyssey Sa Imager and the BioStack plate gripper. (If there is insufficient clearance, the
extended Odyssey Sa Imager plate carrier will collide with the BioStack accessory.)
13) Push the eject button on the Odyssey Sa Imager front panel to extend the plate carrier. Place the
leveling and alignment fixture into the plate carrier as shown below.