6
Switch back to the N
2
O span gas.
7
Wait for the reading to stabilize,.
Under
Options > Calibration > N
2
O Span
, enter the concentration of the gas and
click
Span N
2
O
.
Recovering from a bad calibration
In the event that an attempt to zero or span fails to work as intended, you can
restore previous zero and span settings by selecting the file and clicking
Restore
.
Factory Reset
reloads the factory original zero and span calibration parameters.
Section 6. Maintenance
6-3
Recovering from a bad calibration