Electro-Magnetic Immunity
This EUT is intended for use in the electromagnetic environment specified below.
The customer or the user of the EUT should assure that it is used in such an environment.
Immunity Test
IEC 60601-1-2
Test Level
Compliance Level
Electromagnetic Environment – Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6kV Contact
±8kV Air
±6kV Contact
±8kV AIR
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast transient/
burst
IEC 61000-4-4
±2kV for power
supply lines
± 1kV for input/
output lines
±2kV for power
supply lines
± 1kV for input/
output lines
Mains power quality should be that of a
typical commercial or hospital environment.
Surge
IEC 61000-4-5
±1kV differential
mode
±2kV common mode
±1kV differential
mode
±2kV common
mode
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines)
IEC 61000-4-11
<5% UT
(>95% dip in UT for
0.5 cycle).
40% UT
(60% dip in UT for
5 cycle)
70% UT
(30% dip in UT for
25cycle)
<5% UT
(<95% dip in UT)
for 5s
<5% UT
(>95% dip in UT for
0.5 cycle).
40% UT
(60% dip in UT for
5 cycle)
70% UT
(30% dip in UT for
25cycle)
<5% UT
(<95% dip in UT)
for 5s
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the EUT image intensifier
requires continued operation during power
mains interruptions, it is recommended that
the EUT image intensifier be powered from
an uninterruptible power supply or a battery.
Power frequency
(50/60Hz) magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.
NOTE
•
UT is THE A.C mains voltage prior to application of the test level.