Display and control panel AMS 384i
40
AMS 384
i
Leuze electronic GmbH + Co. KG
LSR
Warning - laser prefailure message:
Laser diode old, device still functional, exchange or have repaired.
TMP
Warning - temperature monitoring:
Internal device temperature above/below permissible range.
PLB
Plausibility error:
Implausible measurement value. Possible causes: light beam interruption, outside
of measurement range, permissible internal device temperature considerably
exceeded or traverse rate >10m/s.
Depending on the configuration, either zero or the last valid measurement value is
output at the interfaces.
ATT
Warning - received signal:
Laser exit window or reflector soiled or fogged by rain, water vapor or fog. Clean or
dry surfaces.
ERR
Internal hardware error:
The device must be sent in for inspection.
Bar graph
Indicates the strength of the received laser light.
The center bar represents the ATT warning threshold. The distance value remains
valid and is output at the interfaces.
If no bar graph is available, the PLB status information appears at the same time.
The measurement value is assessed as implausible. Depending on the configuration,
either zero or the last valid measurement value is output at the interfaces.
Interface info
The abbreviation "IBS" stands for the Interbus interface.
Position value
The measured position value is displayed in the configured unit of measurement.
+87.000m
With the metric setting, the measurement value is always displayed in meters
to three decimal places.
+87.0in
With the inch setting, the measurement value is always displayed in inches to
one decimal place.
Interbus interface
Position value
IO1 LSR PLB
IBS
IO2 TMP ATT
ERR
+ 87.000m