Description ODSL 30
Leuze electronic GmbH + Co. KG
ODSL 30
26
3.6.3 Adjustment of the display contrast
While switching the device on, press both arrow keys of the ODSL 30 simultaneously.
After releasing the keys, you can decrease or increase the contrast of the LC display with the arrow keys
(value range 0 … 255). By pressing ENTER, the adjusted contrast value is applied and you get to the
configuration menu of the ODSL 30.
3.6.4 Reset to factory settings
By pressing ENTER while switching the device on, you can reset the configuration of the ODSL 30 to the
factory settings.
A safety prompt appears.
By pressing ENTER again, all parameters are reset to factory settings. All settings made previously are
permanently lost. By pressing an arrow key, the ODSL 30 returns to measurement operation without reset
-
ting the parameters.
3.6.5 Querying the device software version
You can query the device software version in the menu for configuring the ODSL 30. To do this, select the
following menu item in the
Service Menu
:
3.6.6 Referencing the device
The ODSL 30 is equipped with a referencing function for internally calibrating the sensor.
By carrying out the integrated referencing function before a measurement, the sensor's accuracy can be
improved.
A referencing operation is performed
• when switching on the device (Power-On).
• by means of a signal at the activation/referencing input (PIN 2).
• by means of a command in remote control operation (ODSL 30/D… only).
While the referencing function is carried out (duration approx. 350ms), no measurement can be taken.
contrast: 160
Default Setting?
Press
for OK
SW V01.20 YYMMDD
Val:
31024
<- Software version V0x.xx with date (YY = year, MM = month, DD = day)
NOTE
In particular, the referencing function should be performed for changing environmental condi
-
tions.